1 A Substitution Method for Antenna Calibration by the Use of Broadband Antenna (30 to 1000 MHz) Mitsunobu Samoto 1, Nobuhito Samoto 1, Hiroyuki Shimanoe 1, Ikuo Makino 1 and Kazuo Shimada 2, (1)Liberty Labs Asia, Inc., Yokohama, Japan, (2)ETS Lindgren Japan, Tokyo, Japan 1/ 37 ABSTRACT Calibration of Broadband antennas ANSI C [1] provides calibration method with the Standard Site Method (SSM) at a Standard Antenna Calibration Site (SACS) and two annexes, Annex G and Annex H, for the use of antenna factors for NSA measurements. (47CFR FCC Part 15 Sec. 31) This paper proposes an alternative calibration method to the Annex G and H, that is, a Broadband Antenna Substitution Method (SUB) in 10 m semianechoic chambers (SAC). 2/ 37 要約 - ANSI C [1] は標準アンテナ校正サイト SACS /OATS での標準サイト法 (SSM) による校正 法および NSA 測定 に Annex Annex G と H でアンテナファクタを提供しています この論 は ANSI C の Annex G と H に代わるブロードバンドアンテナの校正を 10 m 半無響暗室 (SAC) 内で う ブロードバンドアンテナ置換 ( 校正 ) 法 (SUB) を提案するものである

2 I. INTRODUCTION 1 Antenna calibration method in ANSI C : Antenna calibration with SSM shall be performed at SACS SACS: The site shall be void of buildings, electric lines, fences, trees, underground cable, pipelines, etc. as specified in ANSI C Characteristics of SACS: Ideal site for the Antenna calibration as specified in CISPR Ed.1.1 CALTS SSM: Most common Calibration Method for antenna calibration specified in ANSI C Paragraph 5, Detail procedures are provided in Annex G and Annex H. GSCF: Correction factors (in db) that are calculated or measured for each frequency at a specific geometry. 3/ 37 I. 前書き-1 ANSI C におけるアンテナ校正 法 SSM によるアンテナ校正は SACS で われなければならない SACS: サイトは ANSI C に規定されているように 建物 電 線 樹 フェンス 地下ケーブル パイプライン等を避けなければならない サイト特性 : アンテナ校正の理想サイトは CISPR C ED 1.1 CALTS として規定されている SSM: アンテナ校正の最も 般的な校正 法は ANSI C 節及び詳細 順は Annex G と H で提供されている GSCF: 指定された特定の配置に対し各周波数について算出または測定された補正係数 (db)

3 I. INTRODUCTION 2 Substitution method (SUB) for the broadband antenna calibration in this paper. Use Reference antennas that are calibrated at the Reference OATS Reference OATS shall be SACS to calibrate Reference antenna Substitute the DUT with Reference antenna to obtain Antenna Factors. SUB can be performed in the 10 meter Semi Anechoic Chamber or OATS for EMI measurements. Measurement Uncertainty degrades at the worst value of only 0.2 db 4/ 37 I. 前書き-2 この論 におけるブロードバンドアンテナによる置換法 (SUB) リファレンス OATS で校正されたリファレンスアンテナを使 する リファレンスアンテナを校正するためのリファレンス OATS は SACS でなければならない アンテナファクタを求めるためには DUT とリファレンスアンテナを置き換える 置換法 (SUB) は 10m 半無響暗室か EMI 測定のための OATS で実施可能である 測定の不確かさ (MU) は最悪でもたった 0.2 db 低下するだけである

4 II. MEASUREMENT CONDITIONs A. Using Broadband antennas Each antenna used in this study is calibrated by the SSM at the reference OATS meeting the requirements of [3] and [4] Biconical Antenna (Bicon): 30 to 300 MHz BBA9106 Bicon 1, Bicon 2 and Bicon 3 Logperiodic Antenna (LPD): 200 to 1000 MHz VULP9118A LPD 1, LPD 2 and LPD 3 Hybrid Antenna (Hybrid): 30 to 1000 MHz VULB9160 Hybrid 1, Hybrid 2 and Hybrid 3 Transmitting Antenna: 30 to 1000 MHz VULB9160 Hybrid Tx (The antenna is validated in house) Network Analyzer: Advantest R3770 5/ 37 II. 測定条件この研究に使 された各アンテナは参照 [3] と [4] の要求を満たすリファレンス OATS で SSM によって校正されています II. MEASUREMENT CONDITIONs A. Using Broadband antennas Each antenna used in this study was calibrated by the SSM at the reference OATS (Ref OATS) meeting the requirements of [3], [4] Ant1 Bicon 1, LPD 1, Hybrid 1 Ant2 Bicon 2, LPD 2, Hybrid 2 NSA Ant3 Bicon 3, LPD 3, Hybrid 3 SSM SUB Reference Antenna DUT Antenna 6/ 37 II. 測定条件この研究に使 された各アンテナは参照 [3] と [4] の要求を満たすリファレンス OATS で SSM によって校正されています

5 II. MEASUREMENT CONDITIONs B. Test sites Reference OATS (Ref OATS): OATS: Meet CALTS requirements of CISPR (2003) [4]. Ground Plane (GP): 50 m 80 m without weather protection enclosure OATS1: OATS GP: 20 m 40 m without weather protection enclosure SAC1: 10 msacsize:24m 15 m height =10 m SAC2: 10 msacsize:23m 14 m height = 9.2 m SAC3: 10 m SAC Size: 18.4 m 9.9 m height = 7.7 m SAC4: 10 m SAC Size: m 14.6 m height = 9.5 m Correction factors to free space (ΔAF) and correction factors for NSA (GSCF) in ANSI C Annex G are not applied in this paper. 7/37 II. 測定条件 III. MEASUREMENT PROCEDUREs A. Broadband Antenna Substitution Method (SUB) Fig. 1. Antenna arrangement of SUB The antenna factor, AF x,ofthedut antenna is calculated as follows; Set a Hybrid at the transmitting (Tx) side and the reference broadband antenna at the receiving (Rx) side as showninfig.1.recordtherxlevel as L ref. Replace the reference broadband antenna at the Rx side to a DUT and record the Rx level as L x. AF x =AF ref +L ref L x (db) (1) where, AF ref areantennafactorsof reference broadband antenna calibrated by the SSM at the Ref OATS. 8/ 37 II. 測定条件 1)DUT のアンテナファクタ AFx は以下のように計算されます 2)Fig.1 のように 送信側にハイブリッドアンテナ 受信側にリファレンスブロードバンドアンテナをセットし 受信レベル Lx を記録します 3) 受信側のリファレンスアンテナを DUT に置き換え 受信レベル Lx を記録します 4)AFx = AFref + Lref Lx (db) (1) 5) ここで AFref はリファレンス OATS で SSM 法によって校正されたリファレンスブロードバンドアンテナのアンテナファクタです

6 III. MEASUREMENT PROCEDUREs B. Standard Site Method (SSM) Fig. 2. Antenna arrangement of SSM Fig. 3. Three site attenuation measurements using three different antennas in pairs According to ANSI C Section 5.2, the SSM requires three site attenuation measurements under identical geometries (h1, h2, R) using three different antennas taken in pairs, as shown in Figure 2. The three equations associated with the three site attenuation measurements are (2), (3), and (4). AF 1 +AF 2 =A 1 +20logf M E D max (2) AF 1 +AF 3 =A 2 +20logf M E D max (3) AF 2 +AF 3 =A 3 +20logf M E D max (4) (All equations in db) AF 1,AF 2 and AF 3 are the antenna factors of antennas 1, 2, and 3 in db (1/m). A 1,A 2 and A 3 are the measured site attenuation in db. (See Figure 3 and Section 5.3 of [1]) 9/ 37 II. 測定条件 ANSI C63.5 セクション 5.2 によれば SSM 法は 3 つの異なるアンテナをペアにして使 し 全く同じジオメトリで 3 回のサイトアッテネーション測定が必要となる 3 つのサイトアッテネーション測定にかかわる 3 程式は (1) (2) と (3) である AF1 AF2 AF3 はアンテナ1,2,3のアンテナファクタである A1 A2 A3 は測定されたサイトアッテネーションである

7 IV. COMPARISON BETWEEN ANTENNA FACTORs OF BROADBAND ANTENNAs MEASURED AT OATS AND SACs BY SSM SSM measurement studies in OATS and SAC AF x at the Ref OATS for 10 m distance, horizontal polarization and 2 m Tx height. AF differences at each site is calculated as follows and shown in Figs. 4, 6 and 8. ΔdB1 = (AF x measured by SSM at each test site) (AF ref by SSM at Ref OATS) (8) The NSA at each site for 10 m distance, 2 m Tx height and horizontal polarization using antennas of Bicon 1 and Bicon 2,LPD 1 and LPD 2, and Hybrid 1 and Hybrid 2. ΔdB2 are calculated as follows and shown Figs. 5, 7 and 9. ΔdB2 = (NSA measured at each test site) (Theoretical NSA) (9) 10 / 37 IV. SSM 法による OATS と SAC におけるブロードバンドアンテナのアンテナファクタの 較 (Bicon) OATS と SAC による SSM 法の検証リファレンス OATS による 10m 距離 Tx さ 2m 平の AF x 各サイトにおける AF 差は以下のように計算され Fig.4,6,8 に されています ΔdB1 = (AF x measured by SSM at each test site) (AF ref by SSM at Ref OATS)(8) 各サイトにおける 10m 距離 Tx さ 2m 平 バイコン LPD ハイブリッドの NSA ΔdB2 は以下のように計算され Fig.5,7,9 に されています ΔdB2 = (NSA measured at each test site) (Theoretical NSA) (9)

8 IV. COMPARISON BETWEEN ANTENNA FACTORs OF BROADBAND ANTENNAs MEASURED AT OATS AND SACs BY SSM (Bicon) Fig. 4. AF Deviation: D10H2 ΔdB1 of Bicon 2 measured at each site Fig. 5. NSA Deviation: D10H2 ΔdB2: Pair of Bicon 1 and Bicon 2 Fig. 4 shows AF differences of ΔdB1 for Bicon 2. Fig. 5 shows the differences of ΔdB2 each test site. Vertical 11 /37 IV. SSM 法による OATS と SAC におけるブロードバンドアンテナのアンテナファクタの 較 (Bicon) Fig.4 は Bicon 2 のΔdB1 のアンテナファクタ偏差を しています Fig.5 は各サイトのΔdB2の偏差を しています IV. COMPARISON BETWEEN ANTENNA FACTORs OF BROADBAND ANTENNAs MEASURED AT OATS AND SACs BY SSM (LPD) Fig. 6. AF Deviation: D10H2 ΔdB1 of LPD 2 measured at each site Fig. 7. NSA Deviation: D10H2 ΔdB2: Pair of LPD 1 and LPD 2 Fig. 6 shows AF differences of ΔdB1 for LPD 2. Fig. 7 shows the differences of ΔdB2 at each test site. Vertical 12 / 37 IV. SSM 法による OATS と SAC におけるブロードバンドアンテナのアンテナファクタの 較 (LPD) Fig.6 は LPD 2 のΔdB1 のアンテナファクタ偏差を しています Fig.7 は各サイトのΔdB2の偏差を しています

9 IV. COMPARISON BETWEEN ANTENNA FACTORs OF BROADBAND ANTENNAs MEASURED AT OATS AND SACs BY SSM (Hybrid) Fig. 8. AF Deviation: D10H2 ΔdB1 of Hybrid 2 measured at each site Fig. 9. NSA Deviation: D10H2 ΔdB2: Pair of Hybrid 1 and Hybrid 2 Fig. 8 shows AF differences of ΔdB1 for Hybrid 2. Fig. 9 shows the differences of ΔdB2 at each test site. Vertical 13 /37 IV. SSM 法による OATS と SAC におけるブロードバンドアンテナのアンテナファクタの 較 (Hybrid) Fig.8 は Hybrid 2 のΔdB1 のアンテナファクタ偏差を しています Fig.9 は各サイトのΔdB2の偏差を しています IV. COMPARISON BETWEEN ANTENNA FACTORs OF BROADBAND ANTENNAs MEASURED AT OATS AND SACs BY SSM AscanbeseenfromcomparisonsofFigs.4and5,Figs.6and7, and Figs. 8 and 9, ΔdB1 characteristics are similar to ΔdB2 characteristics, and deviations of ΔdB1 are one half of ΔdB2, approximately. From these measurement results, it can be seen that the antenna calibration with SSM in SAC is significantly affected by NSA characteristics of the SAC used. 14 / 37 IV. SSM 法による OATS と SAC におけるブロードバンドアンテナのアンテナファクタの 較 Fig. 4 と 5 Fig. 6 と 7 Fig. 8 と 9 の 較から られるように ΔdB1 特性はΔdB2 に似ており そして偏差はおおよそ ΔdB2 の半分が ΔdB1 となっている これらの測定結果から SAC での SSM 法によるアンテナ校正は 使 された SAC の NSA 特性が著しく影響する

10 V. COMPARISON BETWEEN AFs MEASURED AT BOTH OATS AND SACs BY SUB The SUB using a broadband antenna calibrated by the SSM at the Ref OATS as the reference antenna was studied and the specific results are reported (Supplemental data of D10V1 referred) Conditions; Distance of antennas: 10 meters Polarization: Horizontal Height of antennas: 2 m. Reference antennas: Bicon 1,LPD 1 and Hybrid 1 DUT antennas: Bicon 2,LPD 2 and Hybrid 2 15 /37 V. SUB 法による OATS と SAC での測定における AFs の 較リファレンス OATS において SSM 法で校正されたブロードバンドアンテナをリファレンスアンテナとして使 する SUB 法を検証し 具体的な結果を報告します (D10V1 の補 データを参照のこと )

11 V. COMPARISON BETWEEN AFs MEASURED AT BOTH OATS AND SACs BY SUB The differences between AFs measured by the SUB at each test site and AF ref are shown in Figs. 10 to 12 for each antenna of Bicon, LPD and Hybrid as ΔdB3. ΔdB3 = (AFx measured by SUB at each test site) (AFref by SSM at Ref OATS) (10) As can be seen from Figs. 10 to 12, AFs by this SUB coincide within 0.4 db to the AF ref. AFs by the SUB at the OATS are worse than at SAC showing consistent accuracy within / 0.3 db. Those AFs measured by the SUB at SACs would not be affected by NSA characteristics of test sites. The large deviations at the OATS are caused by changes of the measurement condition such as wind breezing and temperature changes. 16 / 37 V. SUB 法による OATS と SAC での測定における AFs の 較 バイコン LPD ハイブリッドを各サイトで SUB 法によって校正された AF s とAF ref の差分 ΔdB3 として Fig.10 から 12 で す ΔdB3 = (AF x measured by SUB at each test site) (AF ref by SSM at Ref OATS) (10) Fig.10 から 12 で分かるように SUB 法による AFs は 0.4dB 以内で AF ref と 致する OATS での SUB 法による AF s は SAC での +0.2/-0.3dB という 貫した精度よりは悪い SAC で SUB 法にて測定されるそれらの AF s はテストサイトの NSA 特性の影響を受けない OATS における きな偏差の原因は気温変化や による揺れといった測定条件の変化によるものである

12 V. COMPARISON BETWEEN AFs MEASURED AT BOTH OATS AND SACs BY SUB (Bicon) Fig. 10. Differences of AFs (ΔdB3) for Bicon 2 at each site and AF ref Fig. 5. NSA Deviation: D10H2 ΔdB2: Pair of Bicon 1 and Bicon 2 Fluctuations of AFs from the OATS may be caused by wind breezing. Vertical 17 / 37 V. SUB 法による OATS と SAC での測定における AF s の 較 (Bicon) V. COMPARISON BETWEEN AFs MEASURED AT BOTH OATS AND SACs BY SUB (LPD) Fig. 11. Differences of AFs (ΔdB3) for LPD 2 at each site and AF ref Fig. 7. NSA Deviation: D10H2 ΔdB2: Pair of LPD 1 and LPD 2 Fluctuations of AFs from the OATS may be caused by wind breezing. Vertical 18 / 37 V. SUB 法による OATS と SAC での測定における AF s の 較 (LPD)

13 V. COMPARISON BETWEEN AFs MEASURED AT BOTH OATS AND SACs BY SUB (Hybrid) Fig. 12. Differences of AFs (ΔdB3) for Hybrid 2 at each site and AF ref Fig. 9. NSA Deviation: D10H2 ΔdB2: Pair of Hybrid 1 and Hybrid 2 Fluctuations of AFs from the OATS may be caused by wind breezing. Vertical 19 / 37 V. SUB 法による OATS と SAC での測定における AF s の 較 (Hybrid) VI. CAUTIONs OF SETUP AND MEASUREMENT UNCERTAINTY FOR SUB AT SACs A. Precautions The following cautions must be carefully kept by procedures instructions; B. Uncertainty Estimation for Broadband Antenna Calibration by SUB Heat up time Consistent reference antenna calibration Distance and height of antennas Cable treatment 20 / 37 VI. SAC における SUB 法のための測定不確かさの為の注意事項 A. 事前注意以下の注意事項は 順指 に注意深く従うこと ヒートアップ時間 貫したリファレンスアンテナ校正 アンテナ距離と さ ケーブルの取り扱い

14 VI. CAUTIONs OF SETUP AND MEASUREMENT UNCERTAINTY FOR SUB AT SACs C. Considerations to Measurement Uncertainty(MU) on Antenna Calibration No. 1 & 2: Measurement Uncertainty of Calibration factor of Reference antennas for SUB. MU for AF ref for SUB; Bicon: 0.96 db, LPD: 0.6 db, Hybrid: 0.68 db No. 3 & 4: Amplitude resolution and Dynamic Accuracy of Network Analyzer s Specification No. 5: Cables characteristics No. 6 & 7: Influence by Electrical Uniformity No. 8: Distance setting errors. No. 9: The height setting errors to DUT antenna. No. 10: Vertical face alignment errors of both antennas of Bicon and LPA. No. 11: Error of setting distance between antenna end and antenna mast. No. 12 & 13: This error is estimated from 5 times measurements No. 14 & 15: Mismatch loss Tx side has no influence by SUB. 21 / 37 VI. SAC における SUB 法のための測定不確かさの為の注意事項 No.1&2: No.3&4: No.5: No.6&7: No.8: No.9: No.10: No.11: No.12&13: No.14&15: SUB 法の為のリファレンスアンテナの校正ファクタの測定不確かさネットワークアナライザの振幅分解能とダイナミック精度ケーブル特性電界均 性の影響距離のセッティングエラー DUT アンテナの さのセッティングエラーバイコンと LPD の垂直 のアレンジエラーアンテナマストとアンテナ端の距離のセッティングエラー 5 回測定によって計算されるエラー SUB 法による Tx 側のミスマッチは影響ありません

15 VII. CONCLUSION Antenna calibration with SSM shall use a Standard Antenna Calibration Site (SACS)/ ANSI C _5.2 and should not be applied in the 10 m semi anechoic chambers (SAC) especially for Bicon and Bilog antennas. Advantage on antenna calibration with SUB: Consistent Calibration: less affects from ambient conditions of wind, rain, radio noises and NSA characteristics. Time saving: AF measurements by the SUB take about two third less than that of SSM. Disadvantage on antenna calibration with SUB: Measurement Uncertainty will be degraded around 0.2dB to that of Ref OATS at the worst case. SUB will be acceptable for the broadband antenna calibration method. 22 / 37 VII. まとめ SSM 法によるアンテナ校正は SACS( 基準アンテナ校正サイト ) を使 し 特にバイコン及びハイブリッドアンテナにおいては 10m SAC は適 されるべきではない SUB 法によるアンテナ校正の利点 貫性のある校正 : 電波ノイズといった環境条件や NSA 特性の影響が少ない 時間短縮 :SUB 法による AF 測定は SSM 法の 2/3 で済む SUB 法によるアンテナ校正の 点 測定不確かさが悪くてリファレンス OATS のより 0.2 程度悪くなる SUB 法はブロードバンドアンテナの校正 法として適している

16 REFERENCES [1] ANSI C : American National Standard Electromagnetic Compatibility Radiated Emission Measurements in Electromagnetic Interference (EMI) Control Calibration of Antennas (9 khz to 40 GHz)(Revision of ANSI C ) [2] ANSI C : American National Standard for Methods of Measurement of Radio Noise Emissions from Low Voltage Electrical and Electronic Equipment in the Range of 9 khz to 40 GHz (Revision of ANSI C ) [3] ANSI C : American National Standard Guide for Construction of Open Area Test Sites for Performing Radiated Emission Measurements (Revision of ANSI C ) [4] CISPR ( ): Specification for radio disturbance and immunity measuring apparatus and methods Part 1 5: Radio disturbance and immunity measuring apparatus Antenna calibration test sites for 30 MHz to MHz 23 / 37 Thank you 24 / 37

17 Supplemental data 1 Reference OATS (Ref OATS) Meet CALTS requirements of CISPR (2003) [4]. Ground Plane (GP): 50 m 80 m without weather protection enclosure 25 / 37 Supplemental data 2 Test sites: OATS1 OATS GP: 20 m 40 m without weather protection enclosure 26 / 37 Supplemental data 3 Test sites: SAC1 10 m SAC Size: 24 m 15 m height =10 m 27 / 37

18 Supplemental data 4 Test sites: SAC2 10 m SAC Size: 23 m 14 m height = 9.2 m 28 / 37 Supplemental data 5 Test sites: SAC3 10 m SAC Size: 18.4 m 9.9 m height = 7.7 m 29 / 37 Supplemental data 6 Test sites: SAC4 10 m SAC Size: m 14.6 m height = 9.5 m 30 / 37

19 Supplemental data 7 Uncertainty Estimation for Broadband Antenna Calibration by SUB No Source of error Value (db) Probability Distribution k Ui (y) 1 Uncertainty of ref antenna: MHz 0.96 Normal Uncertainty of ref antenna: MHz 0.68 Normal NA Linearity NA Stability / Reading fluctuation Cable loss fluctuation Uniformity in calibration area: MHz Uniformity in calibration area: MHz Antenna distance: ±2cm / Δ=20log(9.98/10) Antenna height: ±1cm Radiation pattern levelness / facing setting: ± ANT mast influence Measurement repeatability / ANT swing: MHz 0.15 Normal Measurement repeatability / ANT swing: MHz 0.20 Normal Rx mismatch:30 300MHz Rx mismatch: MHz Extended uncertainty (Bicon/Hybrid : MHz) k= Extended uncertainty (LPD/Hybrid : MHz) k= / 37 Supplemental data 8 SSM Distance 10m Vertical 1m Fig. 13. SSM AF Deviation: D10V1 ΔdB1 of Bicon 2 measured at each site Fig. 14. NSA Deviation ΔdB2: Pair of Bicon 1 and Bicon 2 32 / 37 Supplemental data 9 SSM Distance 10m Vertical 1m Fig. 15. SSM AF Deviation: D10V1 ΔdB1 of LPD 2 measured at each site Fig. 16. NSA Deviation: D10V1 ΔdB2: Pair of LPD 1 and LPD 2 33 / 37

20 Supplemental data 10 SSM Distance 10m Vertical 1m Fig. 17. SSM AF Deviation: D10V1 ΔdB1 of Hybrid 2 measured at each site Fig. 18. NSA Deviation: D10V1 ΔdB2: Pair of Hybrid 1 and Hybrid 2 34 / 37 Supplemental data 11 SUB Distance 10m Vertical 1m Fig. 19. SUB Differences AFs (ΔdB3) for Bicon2 at each site and AFref Fig. 14. NSA Deviation ΔdB2: Pair of Bicon 1 and Bicon 2 35 / 37 Supplemental data 12 SUB Distance 10m Vertical 1m Fig. 20. SUB Differences AFs (ΔdB3) for LPD 2 at each site and AF ref Fig. 16. NSA Deviation: D10V1 ΔdB2: Pair of LPD 1 and LPD 2 36 / 37

21 Supplemental data 13 SUB Distance 10m Vertical 1m Fig. 21. SUB Differences AFs (ΔdB3) for Hybrid 2 at each site and AF ref Fig. 18. NSA Deviation: D10V1 ΔdB2: Pair of Hybrid 1 and Hybrid 2 37 / 37

CISPR 32の概要 (ed. 2)

CISPR 32の概要 (ed. 2) CISPR 32 (ed. 2) e 2017 7 25 1 1 2 2 2.1.................. 2 2.2............... 2 2.2.1.......... 3 3 3 3.1.............. 4 3.2 EUT............. 4 3.2.1 EUT.............. 4 3.2.2.............. 4 3.3...............

More information

IEC :2014 (ed. 4) の概要 (ed. 2)

IEC :2014 (ed. 4) の概要 (ed. 2) IEC 60601-1-2:2014 (ed. 4) (ed. 2) e 2018 4 2 1 1 2 / 1 2.1............... 2 2.2............... 3 2.3.................. 4 3 6 4 6 4.1.................. 6 4.1.1............... 7 4.1.2....... 7 4.1.3............

More information

WARNING To reduce the risk of fire or electric shock,do not expose this apparatus to rain or moisture. To avoid electrical shock, do not open the cabi

WARNING To reduce the risk of fire or electric shock,do not expose this apparatus to rain or moisture. To avoid electrical shock, do not open the cabi ES-600P Operating Instructions WARNING To reduce the risk of fire or electric shock,do not expose this apparatus to rain or moisture. To avoid electrical shock, do not open the cabinet. Refer servicing

More information

Tab 5, 11 Tab 4, 10, Tab 3, 9, 15Tab 2, 8, 14 Tab 1, 7, 13 2

Tab 5, 11 Tab 4, 10, Tab 3, 9, 15Tab 2, 8, 14 Tab 1, 7, 13 2 COMPANION 20 MULTIMEDIA SPEAKER SYSTEM Owner s Guide Tab 5, 11 Tab 4, 10, Tab 3, 9, 15Tab 2, 8, 14 Tab 1, 7, 13 2 Tab1, 7, 13 Tab 2, 8, 14 Tab 3, 9, 15 Tab 4, 10, Tab 5, 11 This product conforms to all

More information

Visual Evaluation of Polka-dot Patterns Yoojin LEE and Nobuko NARUSE * Granduate School of Bunka Women's University, and * Faculty of Fashion Science,

Visual Evaluation of Polka-dot Patterns Yoojin LEE and Nobuko NARUSE * Granduate School of Bunka Women's University, and * Faculty of Fashion Science, Visual Evaluation of Polka-dot Patterns Yoojin LEE and Nobuko NARUSE * Granduate School of Bunka Women's University, and * Faculty of Fashion Science, Bunka Women's University, Shibuya-ku, Tokyo 151-8523

More information

5 11 3 1....1 2. 5...4 (1)...5...6...7...17...22 (2)...70...71...72...77...82 (3)...85...86...87...92...97 (4)...101...102...103...112...117 (5)...121...122...123...125...128 1. 10 Web Web WG 5 4 5 ²

More information

Huawei G6-L22 QSG-V100R001_02

Huawei  G6-L22 QSG-V100R001_02 G6 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 1 2 3 17 4 5 18 UI 100% 8:08 19 100% 8:08 20 100% 8:08 21 100% 8:08 22 100% 8:08 ********** 23 100% 8:08 Happy birthday! 24 S S 25 100% 8:08 26 http://consumer.huawei.com/jp/

More information

Studies of Foot Form for Footwear Design (Part 9) : Characteristics of the Foot Form of Young and Elder Women Based on their Sizes of Ball Joint Girth

Studies of Foot Form for Footwear Design (Part 9) : Characteristics of the Foot Form of Young and Elder Women Based on their Sizes of Ball Joint Girth Studies of Foot Form for Footwear Design (Part 9) : Characteristics of the Foot Form of Young and Elder Women Based on their Sizes of Ball Joint Girth and Foot Breadth Akiko Yamamoto Fukuoka Women's University,

More information

EQUIVALENT TRANSFORMATION TECHNIQUE FOR ISLANDING DETECTION METHODS OF SYNCHRONOUS GENERATOR -REACTIVE POWER PERTURBATION METHODS USING AVR OR SVC- Ju

EQUIVALENT TRANSFORMATION TECHNIQUE FOR ISLANDING DETECTION METHODS OF SYNCHRONOUS GENERATOR -REACTIVE POWER PERTURBATION METHODS USING AVR OR SVC- Ju EQUIVALENT TRANSFORMATION TECHNIQUE FOR ISLANDING DETECTION METHODS OF SYNCHRONOUS GENERATOR -REACTIVE POWER PERTURBATION METHODS USING AVR OR SVC- Jun Motohashi, Member, Takashi Ichinose, Member (Tokyo

More information

Study on Application of the cos a Method to Neutron Stress Measurement Toshihiko SASAKI*3 and Yukio HIROSE Department of Materials Science and Enginee

Study on Application of the cos a Method to Neutron Stress Measurement Toshihiko SASAKI*3 and Yukio HIROSE Department of Materials Science and Enginee Study on Application of the cos a Method to Neutron Stress Measurement Toshihiko SASAKI*3 and Yukio HIROSE Department of Materials Science and Engineering, Kanazawa University, Kakuma-machi, Kanazawa-shi,

More information

untitled

untitled JIS A 12190000 Method for standard penetration test 2005 1 ISO 22476-3:2005 J J 1 SPTSPT 63.5 kg 760 mm SPT N SPT 300 mm N A 2 ISO 22475-1,Geotechnical investigation and testing Sampling methods and groundwater

More information

技術研究報告第26号

技術研究報告第26号 1) 2) 3) 250Hz 500Hz RESEARCH ON THE PHYSICAL VOLUME OF THE DYNAMIC VIBRATION RESPONSE AND THE REDUCTION OF THE FLOOR IMPACT SOUND LEVEL IN FLOORS OF RESIDENTIAL HOUSING Hideo WATANABE *1 This study was

More information

alternating current component and two transient components. Both transient components are direct currents at starting of the motor and are sinusoidal

alternating current component and two transient components. Both transient components are direct currents at starting of the motor and are sinusoidal Inrush Current of Induction Motor on Applying Electric Power by Takao Itoi Abstract The transient currents flow into the windings of the induction motors when electric sources are suddenly applied to the

More information

Corrections of the Results of Airborne Monitoring Surveys by MEXT and Ibaraki Prefecture

Corrections of the Results of Airborne Monitoring Surveys by MEXT and Ibaraki Prefecture August 31, 2011 Corrections of the Results of Airborne Monitoring Surveys by MEXT and Ibaraki Prefecture The results of airborne monitoring survey by MEXT and Ibaraki prefecture released on August 30 contained

More information

A Nutritional Study of Anemia in Pregnancy Hematologic Characteristics in Pregnancy (Part 1) Keizo Shiraki, Fumiko Hisaoka Department of Nutrition, Sc

A Nutritional Study of Anemia in Pregnancy Hematologic Characteristics in Pregnancy (Part 1) Keizo Shiraki, Fumiko Hisaoka Department of Nutrition, Sc A Nutritional Study of Anemia in Pregnancy Hematologic Characteristics in Pregnancy (Part 1) Keizo Shiraki, Fumiko Hisaoka Department of Nutrition, School of Medicine, Tokushima University, Tokushima Fetal

More information

On the Wireless Beam of Short Electric Waves. (VII) (A New Electric Wave Projector.) By S. UDA, Member (Tohoku Imperial University.) Abstract. A new e

On the Wireless Beam of Short Electric Waves. (VII) (A New Electric Wave Projector.) By S. UDA, Member (Tohoku Imperial University.) Abstract. A new e On the Wireless Beam of Short Electric Waves. (VII) (A New Electric Wave Projector.) By S. UDA, Member (Tohoku Imperial University.) Abstract. A new electric wave projector is proposed in this paper. The

More information

24 Depth scaling of binocular stereopsis by observer s own movements

24 Depth scaling of binocular stereopsis by observer s own movements 24 Depth scaling of binocular stereopsis by observer s own movements 1130313 2013 3 1 3D 3D 3D 2 2 i Abstract Depth scaling of binocular stereopsis by observer s own movements It will become more usual

More information

7,, i

7,, i 23 Research of the authentication method on the two dimensional code 1145111 2012 2 13 7,, i Abstract Research of the authentication method on the two dimensional code Karita Koichiro Recently, the two

More information

Description

Description Metal Hybrid Inductor Description Metal Hybrid Inductor Magnetically shielded Suitable for Large Current Size: 4.3 x 4.3 x H2.1 mm Max. Product weight:.18g (Ref.) Halogen Free available Operating temperature

More information

第62巻 第1号 平成24年4月/石こうを用いた木材ペレット

第62巻 第1号 平成24年4月/石こうを用いた木材ペレット Bulletin of Japan Association for Fire Science and Engineering Vol. 62. No. 1 (2012) Development of Two-Dimensional Simple Simulation Model and Evaluation of Discharge Ability for Water Discharge of Firefighting

More information

, (GPS: Global Positioning Systemg),.,, (LBS: Local Based Services).. GPS,.,. RFID LAN,.,.,.,,,.,..,.,.,,, i

, (GPS: Global Positioning Systemg),.,, (LBS: Local Based Services).. GPS,.,. RFID LAN,.,.,.,,,.,..,.,.,,, i 25 Estimation scheme of indoor positioning using difference of times which chirp signals arrive 114348 214 3 6 , (GPS: Global Positioning Systemg),.,, (LBS: Local Based Services).. GPS,.,. RFID LAN,.,.,.,,,.,..,.,.,,,

More information

THE INSTITUTE OF ELECTRONICS, INFORMATION AND COMMUNICATION ENGINEERS TECHNICAL REPORT OF IEICE.

THE INSTITUTE OF ELECTRONICS, INFORMATION AND COMMUNICATION ENGINEERS TECHNICAL REPORT OF IEICE. THE INSTITUTE OF ELECTRONICS, INFORMATION AND COMMUNICATION ENGINEERS TECHNICAL REPORT OF IEICE. E-mail: {ytamura,takai,tkato,tm}@vision.kuee.kyoto-u.ac.jp Abstract Current Wave Pattern Analysis for Anomaly

More information

n 2 n (Dynamic Programming : DP) (Genetic Algorithm : GA) 2 i

n 2 n (Dynamic Programming : DP) (Genetic Algorithm : GA) 2 i 15 Comparison and Evaluation of Dynamic Programming and Genetic Algorithm for a Knapsack Problem 1040277 2004 2 25 n 2 n (Dynamic Programming : DP) (Genetic Algorithm : GA) 2 i Abstract Comparison and

More information

MIDI_IO.book

MIDI_IO.book MIDI I/O t Copyright This guide is copyrighted 2002 by Digidesign, a division of Avid Technology, Inc. (hereafter Digidesign ), with all rights reserved. Under copyright laws, this guide may not be duplicated

More information

Vol. 51 No (2000) Thermo-Physiological Responses of the Foot under C Thermal Conditions Fusako IWASAKI, Yuri NANAMEKI,* Tomoko KOSHIB

Vol. 51 No (2000) Thermo-Physiological Responses of the Foot under C Thermal Conditions Fusako IWASAKI, Yuri NANAMEKI,* Tomoko KOSHIB Vol. 51 No. 7 587-593 (2000) Thermo-Physiological Responses of the Foot under 22-34 C Thermal Conditions Fusako IWASAKI, Yuri NANAMEKI,* Tomoko KOSHIBA and Teruko TAMURA * Junior College Division, Bunka

More information

Jan THE JAPANESE JOURNAL OF ANTIBIOTICS XL-1 Table 1. Outline of administering doses, routes and sampling times *: 4 ml/hr/kg Bacillus subtilis

Jan THE JAPANESE JOURNAL OF ANTIBIOTICS XL-1 Table 1. Outline of administering doses, routes and sampling times *: 4 ml/hr/kg Bacillus subtilis THE JAPANESE JOURNAL OF ANTIBIOTICS XL-1 Jan. 1987 Jan. 1987 THE JAPANESE JOURNAL OF ANTIBIOTICS XL-1 Table 1. Outline of administering doses, routes and sampling times *: 4 ml/hr/kg Bacillus subtilis

More information

A Study on Food Cost Viewed in Relation to Nutrition (Part 1) Fluctuations in Food Costs by Month Yasuko IZUSHI Kimiko MATSUDA No satisfactory way had been found to apprehend the food cost from the viewpoint

More information

perature was about 2.5 Ž higher than that of the control irrespective of wind speed. With increasing wind speeds of more than 1m/s, the leaf temperatu

perature was about 2.5 Ž higher than that of the control irrespective of wind speed. With increasing wind speeds of more than 1m/s, the leaf temperatu Studies on the Row Covering Methods of Vinylon Cheesecloth to Prevent Cold Injury in Cabbage Daizou IGARASHI*, Masumi OKADA** and Keiichl NAKAYAMA*** *Miura Branch, Kanagawa Horticultural Experiment Station,

More information

IPSJ SIG Technical Report Vol.2016-CE-137 No /12/ e β /α α β β / α A judgment method of difficulty of task for a learner using simple

IPSJ SIG Technical Report Vol.2016-CE-137 No /12/ e β /α α β β / α A judgment method of difficulty of task for a learner using simple 1 2 3 4 5 e β /α α β β / α A judgment method of difficulty of task for a learner using simple electroencephalograph Katsuyuki Umezawa 1 Takashi Ishida 2 Tomohiko Saito 3 Makoto Nakazawa 4 Shigeichi Hirasawa

More information

™…

™… Review The Secret to Healthy Long Life Decrease in Oxidative and Mental Stress My motto is Health is not all. But nothing can be done without health. Health is the most important requisite for all human

More information

2015 ( 27 ) RFID RF RFID, 2., 3., 4. i

2015 ( 27 ) RFID RF RFID, 2., 3., 4. i 2015 ( 27 ) 71143204 2015 ( 27 ) RFID RF 1 100 98 20 20 1 19 1. RFID, 2., 3., 4. i Abstract of Bachelor s Thesis Academic Year 2015 Smart Management System of Rechargeable Batteries with Automatic Identification

More information

Therapy for Asthenopia in Cases of Convergence Insufficiency Hiroko TAKASAKI, C.O.J., Nobuko INAGAMI, C.O.J., and Kayoko TAKENAWA, C.O.J.. Orthoptic c

Therapy for Asthenopia in Cases of Convergence Insufficiency Hiroko TAKASAKI, C.O.J., Nobuko INAGAMI, C.O.J., and Kayoko TAKENAWA, C.O.J.. Orthoptic c Therapy for Asthenopia in Cases of Convergence Insufficiency Hiroko TAKASAKI, C.O.J., Nobuko INAGAMI, C.O.J., and Kayoko TAKENAWA, C.O.J.. Orthoptic clinic, Department of Ophthalmology, (Director: Prof

More information

チョークコイル・リアクタ

チョークコイル・リアクタ THIN TRNSORMRS (MPT TYP) OMSTI(JPN) U S I TYP MPT MPT series are available for the thinner-sized sets and suit to I standard. H MPT I KS5535 KS6435 KS7235 KS7250 KS7270 KS9640 KS9660 KS9680 KS00 75.0 67.0

More information

Fig. 3 Coordinate system and notation Fig. 1 The hydrodynamic force and wave measured system Fig. 2 Apparatus of model testing

Fig. 3 Coordinate system and notation Fig. 1 The hydrodynamic force and wave measured system Fig. 2 Apparatus of model testing The Hydrodynamic Force Acting on the Ship in a Following Sea (1 St Report) Summary by Yutaka Terao, Member Broaching phenomena are most likely to occur in a following sea to relative small and fast craft

More information

Fig. 3 Flow diagram of image processing. Black rectangle in the photo indicates the processing area (128 x 32 pixels).

Fig. 3 Flow diagram of image processing. Black rectangle in the photo indicates the processing area (128 x 32 pixels). Fig. 1 The scheme of glottal area as a function of time Fig. 3 Flow diagram of image processing. Black rectangle in the photo indicates the processing area (128 x 32 pixels). Fig, 4 Parametric representation

More information

Table 1. Assumed performance of a water electrol ysis plant. Fig. 1. Structure of a proposed power generation system utilizing waste heat from factori

Table 1. Assumed performance of a water electrol ysis plant. Fig. 1. Structure of a proposed power generation system utilizing waste heat from factori Proposal and Characteristics Evaluation of a Power Generation System Utilizing Waste Heat from Factories for Load Leveling Pyong Sik Pak, Member, Takashi Arima, Non-member (Osaka University) In this paper,

More information

Influence of Material and Thickness of the Specimen to Stress Separation of an Infrared Stress Image Kenji MACHIDA The thickness dependency of the temperature image obtained by an infrared thermography

More information

A Feasibility Study of Direct-Mapping-Type Parallel Processing Method to Solve Linear Equations in Load Flow Calculations Hiroaki Inayoshi, Non-member

A Feasibility Study of Direct-Mapping-Type Parallel Processing Method to Solve Linear Equations in Load Flow Calculations Hiroaki Inayoshi, Non-member A Feasibility Study of Direct-Mapping-Type Parallel Processing Method to Solve Linear Equations in Load Flow Calculations Hiroaki Inayoshi, Non-member (University of Tsukuba), Yasuharu Ohsawa, Member (Kobe

More information

Author Workshop 20111124 Henry Cavendish 1731-1810 Biot-Savart 26 (1) (2) (3) (4) (5) (6) Priority Proceeding Impact factor Full paper impact factor Peter Drucker 1890-1971 1903-1989 Title) Abstract

More information

<95DB8C9288E397C389C88A E696E6462>

<95DB8C9288E397C389C88A E696E6462> 2011 Vol.60 No.2 p.138 147 Performance of the Japanese long-term care benefit: An International comparison based on OECD health data Mie MORIKAWA[1] Takako TSUTSUI[2] [1]National Institute of Public Health,

More information

Safety Performance of Steel Deck Plate (Flat Decks) Used for Concrete Slab Moulding CONTENTS 1. Introduction ---------------------------------------------------------------- (2) 2. Flat Decks ------------------------------------------------------------------

More information

NotePC 8 10cd=m 2 965cd=m 2 1.2 Note-PC Weber L,M,S { i {

NotePC 8 10cd=m 2 965cd=m 2 1.2 Note-PC Weber L,M,S { i { 12 The eect of a surrounding light to color discrimination 1010425 2001 2 5 NotePC 8 10cd=m 2 965cd=m 2 1.2 Note-PC Weber L,M,S { i { Abstract The eect of a surrounding light to color discrimination Ynka

More information

Motivation and Purpose There is no definition about whether seatbelt anchorage should be fixed or not. We tested the same test conditions except for t

Motivation and Purpose There is no definition about whether seatbelt anchorage should be fixed or not. We tested the same test conditions except for t Review of Seatbelt Anchorage and Dimensions of Test Bench Seat Cushion JASIC Motivation and Purpose There is no definition about whether seatbelt anchorage should be fixed or not. We tested the same test

More information

高周波同軸コネクタ

高周波同軸コネクタ RF circuit ANT Probe Signal in Signal out Probe Signal out Signal in RF circuit ANT Probe Probe Signal in Signal out Signal out Signal in RF ANT. RF ANT. Probe Probe Signal out Signal out Signal in Signal

More information

音響部品アクセサリ本文(AC06)PDF (Page 16)

音響部品アクセサリ本文(AC06)PDF (Page 16) Guide for Electret Condenser Microphones A microphone as an audio-electric converting device, whose audio pickup section has a structure of a condenser consisting of a diaphragm and a back plate opposite

More information

25 Removal of the fricative sounds that occur in the electronic stethoscope

25 Removal of the fricative sounds that occur in the electronic stethoscope 25 Removal of the fricative sounds that occur in the electronic stethoscope 1140311 2014 3 7 ,.,.,.,.,.,.,.,,.,.,.,.,,. i Abstract Removal of the fricative sounds that occur in the electronic stethoscope

More information

Estimation of Photovoltaic Module Temperature Rise Motonobu Yukawa, Member, Masahisa Asaoka, Non-member (Mitsubishi Electric Corp.) Keigi Takahara, Me

Estimation of Photovoltaic Module Temperature Rise Motonobu Yukawa, Member, Masahisa Asaoka, Non-member (Mitsubishi Electric Corp.) Keigi Takahara, Me Estimation of Photovoltaic Module Temperature Rise Motonobu Yukawa, Member, Masahisa Asaoka, Non-member (Mitsubishi Electric Corp.) Keigi Takahara, Member (Okinawa Electric Power Co.,Inc.) Toshimitsu Ohshiro,

More information

16_.....E...._.I.v2006

16_.....E...._.I.v2006 55 1 18 Bull. Nara Univ. Educ., Vol. 55, No.1 (Cult. & Soc.), 2006 165 2002 * 18 Collaboration Between a School Athletic Club and a Community Sports Club A Case Study of SOLESTRELLA NARA 2002 Rie TAKAMURA

More information

6 4 4 9RERE6RE 5 5 6 7 8 9 4 5 6 4 4 5 6 8 4 46 5 7 54 58 60 6 69 7 8 0 9 9 79 0 4 0 0 4 4 60 6 9 4 6 46 5 4 4 5 4 4 7 44 44 6 44 8 44 46 44 44 4 44 0 4 4 5 4 8 6 0 4 0 4 4 5 45 4 5 50 4 58 60 57 54

More information

The Evaluation on Impact Strength of Structural Elements by Means of Drop Weight Test Elastic Response and Elastic Limit by Hiroshi Maenaka, Member Sh

The Evaluation on Impact Strength of Structural Elements by Means of Drop Weight Test Elastic Response and Elastic Limit by Hiroshi Maenaka, Member Sh The Evaluation on Impact Strength of Structural Elements by Means of Drop Weight Test Elastic Response and Elastic Limit by Hiroshi Maenaka, Member Shigeru Kitamura, Member Masaaki Sakuma Genya Aoki, Member

More information

I N S T R U M E N T A T I O N & E L E C T R I C A L E Q U I P M E N T Pressure-resistant gasket type retreat method effective bulk compressibility Fro

I N S T R U M E N T A T I O N & E L E C T R I C A L E Q U I P M E N T Pressure-resistant gasket type retreat method effective bulk compressibility Fro Cable Gland This is the s to use for Cable Wiring in the hazardous location. It is much easier to install and maintenance and modification compared with Conduit Wiring with Sealing Fitting. The Standard

More information

ISO & ISO の概要

ISO & ISO の概要 ISO 7637-2 & ISO 7637-3 e 217 3 13 1 1 2 ISO 7637-2 2 2.1.................... 2 2.1.1 LISN (AN).............. 2 2.1.2 S............... 3 2.1.3.. 3 2.2 ECE Regulation No. 1.5.......... 3 3 ISO 7637-2 4

More information

untitled

untitled JAIS 1 2 1 2 In this paper, we focus on the pauses that partly characterize the utterances of simultaneous interpreters, and attempt to analyze the results of experiments conducted using human subjects

More information

(1 ) (2 ) Table 1. Details of each bar group sheared simultaneously (major shearing unit). 208

(1 ) (2 ) Table 1. Details of each bar group sheared simultaneously (major shearing unit). 208 2463 UDC 621.771.251.09 : 621.791.94: 669.012.5 Improvement in Cold Shear Yield of Bar Mill by Computer Control System Koji INAZAKI, Takashi WASEDA, Michiaki TAKAHASHI, and Toshihiro OKA Synopsis: The

More information

1

1 5-3 Photonic Antennas and its Application to Radio-over-Fiber Wireless Communication Systems LI Keren, MATSUI Toshiaki, and IZUTSU Masayuki In this paper, we presented our recent works on development of

More information

,,.,,.,..,.,,,.,, Aldous,.,,.,,.,,, NPO,,.,,,,,,.,,,,.,,,,..,,,,.,

,,.,,.,..,.,,,.,, Aldous,.,,.,,.,,, NPO,,.,,,,,,.,,,,.,,,,..,,,,., J. of Population Problems. pp.,.,,,.,,..,,..,,,,.,.,,...,.,,..,.,,,. ,,.,,.,..,.,,,.,, Aldous,.,,.,,.,,, NPO,,.,,,,,,.,,,,.,,,,..,,,,., ,,.,,..,,.,.,.,,,,,.,.,.,,,. European Labour Force Survey,,.,,,,,,,

More information

003村江.indd

003村江.indd *1 Study on Room ressure Control at Cleanroom art 1 Experiments on Room ressure Fluctuation with Door Operation and Local Ventilation Operation Yukitada MURAE *1 Tamio IWAMURA *2 Hiroyuki NAGAI *3 Shigeru

More information

塗装深み感の要因解析

塗装深み感の要因解析 17 Analysis of Factors for Paint Depth Feeling Takashi Wada, Mikiko Kawasumi, Taka-aki Suzuki ( ) ( ) ( ) The appearance and quality of objects are controlled by paint coatings on the surfaces of the objects.

More information

論 文 Earnings Management in Pension Accounting and Revised Jones Model Kazuo Yoshida, Nagoya City University 要約本稿では退職給付会計における全ての会計選択を取り上げて 経営者の報告利益管理行動

論 文 Earnings Management in Pension Accounting and Revised Jones Model Kazuo Yoshida, Nagoya City University 要約本稿では退職給付会計における全ての会計選択を取り上げて 経営者の報告利益管理行動 論 文 Earnings Management in Pension Accounting and Revised Jones Model Kazuo Yoshida, Nagoya City University 要約本稿では退職給付会計における全ての会計選択を取り上げて 経営者の報告利益管理行動について包括的な分析を行った 分析の結果 会計基準変更時差異による裁量額が最も大きく 報告利益管理の主要な手段であったことが明らかとなった

More information

1 [1, 2, 3, 4, 5, 8, 9, 10, 12, 15] The Boston Public Schools system, BPS (Deferred Acceptance system, DA) (Top Trading Cycles system, TTC) cf. [13] [

1 [1, 2, 3, 4, 5, 8, 9, 10, 12, 15] The Boston Public Schools system, BPS (Deferred Acceptance system, DA) (Top Trading Cycles system, TTC) cf. [13] [ Vol.2, No.x, April 2015, pp.xx-xx ISSN xxxx-xxxx 2015 4 30 2015 5 25 253-8550 1100 Tel 0467-53-2111( ) Fax 0467-54-3734 http://www.bunkyo.ac.jp/faculty/business/ 1 [1, 2, 3, 4, 5, 8, 9, 10, 12, 15] The

More information

GRANT OF EQUIPMENT AUTHORIZATION Certification Issued Under the Authority of the Federal Communications Commission By: lantronix 7535 Irvine Center Dr

GRANT OF EQUIPMENT AUTHORIZATION Certification Issued Under the Authority of the Federal Communications Commission By: lantronix 7535 Irvine Center Dr GRANT OF EQUIPMENT AUTHORIZATION Certification Issued Under the Authority of the Federal Communications Commission By: lantronix 7535 Irvine Center Drive Suite 100 Irvine, CA 92618 Siemic Inc. 775 Montague

More information

10-渡部芳栄.indd

10-渡部芳栄.indd COE GCOE GP ) b a b ) () ) () () ) ) .. () ) ) ) ) () ........... / / /.... 交付税額 / 経常費 : 右軸交付税額 /( 経常費 授業料 ): 右軸 . ) ()... /.. 自治体負担額 / 交付税額 : 右軸 ()......... / 自治体負担額 / 経常費 : 右軸 - No. - Vol. No. - IDE

More information

1 1 tf-idf tf-idf i

1 1 tf-idf tf-idf i 14 A Method of Article Retrieval Utilizing Characteristics in Newspaper Articles 1055104 2003 1 31 1 1 tf-idf tf-idf i Abstract A Method of Article Retrieval Utilizing Characteristics in Newspaper Articles

More information

Fig, 1. Waveform of the short-circuit current peculiar to a metal. Fig. 2. Waveform of arc short-circuit current. 398 T. IEE Japan, Vol. 113-B, No. 4,

Fig, 1. Waveform of the short-circuit current peculiar to a metal. Fig. 2. Waveform of arc short-circuit current. 398 T. IEE Japan, Vol. 113-B, No. 4, Development of a Quick-Acting Type Fuses for Protection of Low Voltage Distribution Lines Terukazu Sekiguchi, Member, Masayuki Okazaki, Member, Tsuginori Inaba, Member (CRIEPI), Naoki Ikeda, Member, Toshiyuki

More information

- 1 -

- 1 - - 1 - - 2 - - 3 - - 4 - - 5 - - 6 - - 7 - - 8 - - 9 - - 10 - - 11 - - 12 - - 13 - - 14 - - 15 - 1 2 1-16 - 2 3 4 5 6 7-17 - 1 2 1 2 3 4-18 - 1 2 3 4 1 2-19 - 1 2 3 1 2-20 - 3 4 5 6 7 1-21 - 1 2 3 4-22

More information

The Effect of the Circumferential Temperature Change on the Change in the Strain Energy of Carbon Steel during the Rotatory Bending Fatigue Test by Ch

The Effect of the Circumferential Temperature Change on the Change in the Strain Energy of Carbon Steel during the Rotatory Bending Fatigue Test by Ch The Effect of the Circumferential Temperature Change on the Change in the Strain Energy of Carbon Steel during the Rotatory Bending Fatigue Test by Chikara MINAMISAWA, Nozomu AOKI (Department of Mechanical

More information

6 4 45 7ZS 5 59 7 8 94 05 4 5 6 4 5 5 6 8 8 40 45 48 56 60 64 66 66 68 7 78 80 8 7 8 0 0 0 90 0 57 64 69 66 66 69 0 4 4 4 4 4 0 7 48 5 4 4 5 4 4 4 7 46 46 6 46 8 46 48 46 46 4 46 46 4 4 5 4 6 4 9 9 0

More information

三菱電線工業時報 第 105 号 2008 年 10 月 1 High-speed disaster prevention radio information system R-LCX50-4SL-75 LANIEEE b/g R-LCX50-4SL D 2 1

三菱電線工業時報 第 105 号 2008 年 10 月 1 High-speed disaster prevention radio information system R-LCX50-4SL-75 LANIEEE b/g R-LCX50-4SL D 2 1 1 GHz GHz 帯長距離漏洩同軸ケーブルを用いた高速防災無線情報システムの研究開発 High Speed Communication System for Emergencies Using Long-distance GHz Leaky Coaxial Cables 長野県工業技術総合センター 国立大学法人信州大学 株式会社サーキットデザイン 加賀電子株式会社 菱星通信システム株式会社ネットワーク技術部

More information

75 unit: mm Fig. Structure of model three-phase stacked transformer cores (a) Alternate-lap joint (b) Step-lap joint 3 4)

75 unit: mm Fig. Structure of model three-phase stacked transformer cores (a) Alternate-lap joint (b) Step-lap joint 3 4) 3 * 35 (3), 7 Analysis of Local Magnetic Properties and Acoustic Noise in Three-Phase Stacked Transformer Core Model Masayoshi Ishida Kenichi Sadahiro Seiji Okabe 3.7 T 5 Hz..4 3 Synopsis: Methods of local

More information

Keysight MIMO MIMO Cluster n Path n σ n, AoA σ n, AoD Θ n, AoA MS/UE Array Boresight Rx0 Tx0 Θ n, AoD LOS BS Array Boresight Θ n+1, AoA Rx1 Tx1 Path n

Keysight MIMO MIMO Cluster n Path n σ n, AoA σ n, AoD Θ n, AoA MS/UE Array Boresight Rx0 Tx0 Θ n, AoD LOS BS Array Boresight Θ n+1, AoA Rx1 Tx1 Path n Keysight MIMO MIMO Cluster n Path n σ n, AoA σ n, AoD Θ n, AoA MS/UE Array Boresight Rx0 Tx0 Θ n, AoD LOS BS Array Boresight Θ n+1, AoA Rx1 Tx1 Path n+1 Cluster n+1 ... 3 1. MIMO... 3 1.1 MIMO 1.2 MIMO

More information

6 4 45 ZS7ZS4ZS 5 59 7 8 94 05 4 5 6 4 5 5 6 8 8 40 45 48 56 60 64 66 66 68 7 78 80 8 7 8 0 0 0 90 0 0 4 4 4 4 6 57 64 69 66 66 66 69 4 0 7 48 5 4 4 5 4 4 4 7 46 46 6 46 8 46 48 46 46 4 46 46 4 4 5 4

More information

1..FEM FEM 3. 4.

1..FEM FEM 3. 4. 008 stress behavior at the joint of stringer to cross beam of the steel railway bridge 1115117 1..FEM FEM 3. 4. ABSTRACT 1. BackgroundPurpose The occurrence of fatigue crack is reported in the joint of

More information

0801391,繊維学会ファイバ12月号/報文-01-西川

0801391,繊維学会ファイバ12月号/報文-01-西川 Pattern Making Method and Evaluation by Dots of Monochrome Shigekazu Nishikawa 1,MarikoYoshizumi 1,andHajime Miyake 2 1 Miyagi University of Education, 149, Aramaki-aza-Aoba, Aoba-ku, Sendai-shi, Miyagi

More information

JOURNAL OF THE JAPANESE ASSOCIATION FOR PETROLEUM TECHNOLOGY VOL. 66, NO. 6 (Nov., 2001) (Received August 10, 2001; accepted November 9, 2001) Alterna

JOURNAL OF THE JAPANESE ASSOCIATION FOR PETROLEUM TECHNOLOGY VOL. 66, NO. 6 (Nov., 2001) (Received August 10, 2001; accepted November 9, 2001) Alterna JOURNAL OF THE JAPANESE ASSOCIATION FOR PETROLEUM TECHNOLOGY VOL. 66, NO. 6 (Nov., 2001) (Received August 10, 2001; accepted November 9, 2001) Alternative approach using the Monte Carlo simulation to evaluate

More information

m m Satoshi SATO 48

m m Satoshi SATO 48 46 22 3 23 REPORT OF HYDROGRAPHIC AND OCEANOGRAPHIC RESEARCHES No.46 March, 2010 Activities on Tides at Hydrographic Department in Meiji Era Satoshi SATO : Environmental and Oceanographic Division Abstract

More information

untitled

untitled 11-19 2012 1 2 3 30 2 Key words acupuncture insulated needle cervical sympathetick trunk thermography blood flow of the nasal skin Received September 12, 2011; Accepted November 1, 2011 I 1 2 1954 3 564-0034

More information

untitled

untitled 20 * Re-Evaluation of Isoseismal Maps and Magnitudes from Two Big Earthquakes along the Subduction Zone of Kyushu and Ryukyu Islands Early in the 20th Century Masayuki TAKEMURA, Katsuhisa KANDA Kobori

More information

„h‹¤.05.07

„h‹¤.05.07 Japanese Civilian Control in the Cold War Era Takeo MIYAMOTO In European and American democratic countries, the predominance of politics over military, i.e. civilian control, has been assumed as an axiom.

More information

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part Reservdelskatalog MIKASA MT65H vibratorstamp EPOX Maskin AB Postadress Besöksadress Telefon Fax e-post Hemsida Version Box 6060 Landsvägen 1 08-754 71 60 08-754 81 00 info@epox.se www.epox.se 1,0 192 06

More information

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part Reservdelskatalog MIKASA MVB-85 rullvibrator EPOX Maskin AB Postadress Besöksadress Telefon Fax e-post Hemsida Version Box 6060 Landsvägen 1 08-754 71 60 08-754 81 00 info@epox.se www.epox.se 1,0 192 06

More information

INDEX PAGE 1. Evaluation Method 1 1. 測定回路 Measurement Circuits 3 (1) 静特性 待機電力特性 通電ドリフト特性 その他特性 Steady state, Standby power, Warm up voltage drift and

INDEX PAGE 1. Evaluation Method 1 1. 測定回路 Measurement Circuits 3 (1) 静特性 待機電力特性 通電ドリフト特性 その他特性 Steady state, Standby power, Warm up voltage drift and C268 53 1 1/35 INDEX PAGE 1. Evaluation Method 1 1. 測定回路 Measurement Circuits 3 (1) 静特性 待機電力特性 通電ドリフト特性 その他特性 Steady state, Standby power, Warm up voltage drift and Other characteristics (2) 入力サージ電流 (

More information

EVALUATION OF NOCTURNAL PENILE TUMESCENCE (NPT) IN THE DIFFERENTIAL DIAGNOSIS OF IMPOTENCE Masaharu Aoki, Yoshiaki Kumamoto, Kazutomi Mohri and Kazunori Ohno Department of Urology, Sapporo Medical College

More information

(43) Vol.33, No.6(1977) T-239 MUTUAL DIFFUSION AND CHANGE OF THE FINE STRUCTURE OF WET SPUN ANTI-PILLING ACRYLIC FIBER DURING COAGULATION, DRAWING AND

(43) Vol.33, No.6(1977) T-239 MUTUAL DIFFUSION AND CHANGE OF THE FINE STRUCTURE OF WET SPUN ANTI-PILLING ACRYLIC FIBER DURING COAGULATION, DRAWING AND (43) Vol.33, No.6(1977) T-239 MUTUAL DIFFUSION AND CHANGE OF THE FINE STRUCTURE OF WET SPUN ANTI-PILLING ACRYLIC FIBER DURING COAGULATION, DRAWING AND DRYING PROCESSES* By Hiroshi Aotani, Katsumi Yamazaki

More information

06_学術_関節単純X線画像における_1c_梅木様.indd

06_学術_関節単純X線画像における_1c_梅木様.indd Arts and Sciences X The formulation of femoral heard measurement corrected enlargement ratio using hip joints X-ray Imaging 1 2 1 1 1 2 Key words: Bipolar Hip Arthroplasty (BHA) Preoperative planning Enlargement

More information

L C -6D Z3 L C -0D Z3 3 4 5 6 7 8 9 10 11 1 13 14 15 16 17 OIL CLINIC BAR 18 19 POWER TIMER SENSOR 0 3 1 3 1 POWER TIMER SENSOR 3 4 1 POWER TIMER SENSOR 5 11 00 6 7 1 3 4 5 8 9 30 1 3 31 1 3 1 011 1

More information

013858,繊維学会誌ファイバー1月/報文-02-古金谷

013858,繊維学会誌ファイバー1月/報文-02-古金谷 Development of Non-Contact Measuring Method for Final Twist Number of Double Ply Staple Yarn Keizo Koganeya 1, Youichi Yukishita 1, Hirotaka Fujisaki 1, Yasunori Jintoku 2, Hironori Okuno 2, and Motoharu

More information

百人一首かるた選手の競技時の脳の情報処理に関する研究

百人一首かるた選手の競技時の脳の情報処理に関する研究 Mem. Faculty. B. O. S. T. Kinki University No. 24 : 33 43 (2009) 33 34 Memoirs of The Faculty of B. O. S. T. of Kinki University No. 24 (2009) になる 音 の 先 頭 ゆら 平 均 30~ 羽 田 後 に 取 り 札 を 認 識 していることを 見 出 した0

More information

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part Reservdelskatalog MIKASA MVC-50 vibratorplatta EPOX Maskin AB Postadress Besöksadress Telefon Fax e-post Hemsida Version Box 6060 Landsvägen 1 08-754 71 60 08-754 81 00 info@epox.se www.epox.se 1,0 192

More information

Mikio Yamamoto: Dynamical Measurement of the E-effect in Iron-Cobalt Alloys. The AE-effect (change in Young's modulus of elasticity with magnetization

Mikio Yamamoto: Dynamical Measurement of the E-effect in Iron-Cobalt Alloys. The AE-effect (change in Young's modulus of elasticity with magnetization Mikio Yamamoto: Dynamical Measurement of the E-effect in Iron-Cobalt Alloys. The AE-effect (change in Young's modulus of elasticity with magnetization) in the annealed state of iron-cobalt alloys has been

More information

ON A FEW INFLUENCES OF THE DENTAL CARIES IN THE ELEMENTARY SCHOOL PUPIL BY Teruko KASAKURA, Naonobu IWAI, Sachio TAKADA Department of Hygiene, Nippon Dental College (Director: Prof. T. Niwa) The relationship

More information

スライド 1

スライド 1 September 12, 2011 Results of Airborne Monitoring Survey by MEXT in the Western Part of Fukushima Prefecture The results of the airborne monitoring survey by MEXT in Fukushima prefecture (announced on

More information

国土技術政策総合研究所 研究資料

国土技術政策総合研究所 研究資料 ISSN TECHNICAL NOTE of National Institute for Land and Infrastructure Management No256 September 2005 Experimental Study on Seismic Behavior of Seawalls for Controlled Waste Disposal Shingo KANO, Katsuya

More information

A comparison of abdominal versus vaginal hysterectomy for leiomyoma and adenomyosis Kenji ARAHORI, Hisasi KATAYAMA, Suminori NIOKA Department of Obstetrics and Gnecology, National Maizuru Hospital,Kyoto,

More information

T05_Nd-Fe-B磁石.indd

T05_Nd-Fe-B磁石.indd Influence of Intergranular Grain Boundary Phases on Coercivity in Nd-Fe-B-based Magnets Takeshi Nishiuchi Teruo Kohashi Isao Kitagawa Akira Sugawara Hiroyuki Yamamoto To determine how to increase the coercivity

More information

INDEX PAGE. 静電気放電イミュニティ試験 3 Electrostatic Discharge Immunity Test (IEC ) 2. 放射性無線周波数電磁界イミュニティ試験 4 Radiated Radio Frequency Electromagnetic Fiel

INDEX PAGE. 静電気放電イミュニティ試験 3 Electrostatic Discharge Immunity Test (IEC ) 2. 放射性無線周波数電磁界イミュニティ試験 4 Radiated Radio Frequency Electromagnetic Fiel A27 58 0A /7 INDEX PAGE. 静電気放電イミュニティ試験 3 Electrostatic Discharge Immunity Test (IEC6000 4 2) 2. 放射性無線周波数電磁界イミュニティ試験 4 Radiated Radio Frequency Electromagnetic Field Immunity Test (IEC6000 4 3) 3. 電気的ファーストトランジェントバーストイミュニティ試験

More information

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part

How to read the marks and remarks used in this parts book. Section 1 : Explanation of Code Use In MRK Column OO : Interchangeable between the new part Reservdelskatalog MIKASA MCD-L14 asfalt- och betongsåg EPOX Maskin AB Postadress Besöksadress Telefon Fax e-post Hemsida Version Box 6060 Landsvägen 1 08-754 71 60 08-754 81 00 info@epox.se www.epox.se

More information

Abstract This paper concerns with a method of dynamic image cognition. Our image cognition method has two distinguished features. One is that the imag

Abstract This paper concerns with a method of dynamic image cognition. Our image cognition method has two distinguished features. One is that the imag 2004 RGB A STUDY OF RGB COLOR INFORMATION AND ITS APPLICATION 03R3237 Abstract This paper concerns with a method of dynamic image cognition. Our image cognition method has two distinguished features. One

More information

NINJAL Research Papers No.8

NINJAL Research Papers No.8 (NINJAL Research Papers) 8: 177 196 (2014) ISSN: 2186-134X print/2186-1358 online 177 3 3 3 1940 3 late adoption real time 3 apparent time * 1. 1 2 3 1.1 3 1 1953 * 2014 3 18 2014 5 13 109 NINJAL 2012

More information

浜松医科大学紀要

浜松医科大学紀要 On the Statistical Bias Found in the Horse Racing Data (1) Akio NODA Mathematics Abstract: The purpose of the present paper is to report what type of statistical bias the author has found in the horse

More information

untitled

untitled TZ-BDT910M TZ-BDT910F TZ-BDT910P μ μ μ μ TM VQT3F51-1 l l l [HDD] [BD-RE] [BD-R] [DVD-V] [BD-V] [RAM] [CD] [SD] [-R] [USB] [-RW] [RAM AVCREC ] [-R AVCREC ] [RAM VR ][-R VR ] [-RW VR ] [-R V ] [-RW

More information