VarianIMRT IMRT MLC LMC VARIAN ONCOLOGY SYSTEMS
|
|
|
- さあしゃ みょうだに
- 7 years ago
- Views:
Transcription
1 1 VARIAN ONCOLOGY SYSTEMS
2 VarianIMRT IMRT MLC LMC VARIAN ONCOLOGY SYSTEMS
3 Varian装置によるIMRT 品質担保の為に必要な作業 Changes Over Time: ICRU 50 to 62 IRRADIATED VOLUME TREATED VOLUME PTV ITV CTV GTV PRV: Includes margin around the OAR to compensate for changes in shape and internal motion and for set-up variation. OAR IAEA 再現在のある患者固定法の選択 マージンの決定 近畿大学 奥村先生からの資料より VARIAN ONCOLOGY SYSTEMS PRV 必要な輪郭の指示 ICRU50/62 線量制約の統一
4 IMRT Contouring Field Setup (Optional) BAO Optimization Fluence Optimization Objectives Optimal Fluence VARIAN ONCOLOGY SYSTEMS
5 IMRT Optimal Fluence LMC Parameters Leaf Motion Calculation MLC Parameters AAA, PBC, Acuros XB algorithm Actual Fluence 3D dose calculation Leaf Motions Delivery VARIAN ONCOLOGY SYSTEMS
6 IMRT Only b5mm VARIAN ONCOLOGY SYSTEMS
7 IMRT IMRTRapidArc NTO IMRT Relative number (0 1000) Defines importance of the objective VARIAN ONCOLOGY SYSTEMS
8 IMRT Upper Objective Lower Objective Objectives on a Line VARIAN ONCOLOGY SYSTEMS
9 IMRT Objective - Function CF = N i=1 ( u( D min d i ) 2 + o( d i D max ) 2 ) 不足線量での評価 不足線量 過線量の項において指定線量の 範囲外は評価しないようにしておく 過線量での評価但し :i= 指定した腫瘍 N= 腫瘍内のボクセル数 d= 線量! u= アンダードーズの係数! p= 指定線量! Dmin= 腫瘍内最少線量 CF= 評価関数 o= オーバードーズの係数 Dmax= 腫瘍内最大線量 VARIAN ONCOLOGY SYSTEMS
10 IMRT Objective - Function CF = p k N k ( d i,k D max,k ) 2 i=1 臓器毎のペナルティー リスク臓器内のボクセルの線量 許容線量 ( 最大線量 ) 但し :k= 臓器! i= ボクセル数! p= ペナルティー係数! N= 登録臓器数! d= ボクセル線量! Dmax= 最大線量! VARIAN VARIAN ONCOLOGY SYSTEMS!10
11 IMRT Normal Tissue Objective VARIAN ONCOLOGY SYSTEMS
12 IMRT Normal Tissue Objective (NTO) f (r) = $ & % '& f 0 e k(r r start ) + f (1 e k(r r start ) ), r r start f 0, r < r start Where r = Distance from target border r start = start of fall off f 0 = start dose (%) f = end dose (%) k = Fall off VARIAN ONCOLOGY SYSTEMS
13 IMRT Smoothing Objectives XY XVarian MLCMU YMUTongue & Groove VARIAN ONCOLOGY SYSTEMS
14 IMRT Minimize Dose Objective Normal Tissue Objective VARIAN ONCOLOGY SYSTEMS
15 IMRT Optimization Parameters Beamlet (IMRT)! Base dose plan Maximum number of iterations Maximum optimization time VARIAN ONCOLOGY SYSTEMS
16 IMRT Point SamplingCalculation Grid IMRT VARIAN ONCOLOGY SYSTEMS
17 IMRT Point SamplingCalculation Grid VARIAN ONCOLOGY SYSTEMS
18 IMRT Optimization - Volume Representation EIP VARIAN ONCOLOGY SYSTEMS
19 IMRT Optimization - Field Representation Ray trace!!! row = leaf pair! Size of the beamlet Leaf width x 2.5 mm VARIAN ONCOLOGY SYSTEMS
20 IMRT Optimization - Target Masking 0.5cm 0.5cmY VARIAN ONCOLOGY SYSTEMS
21 IMRT Optimization - Dose Computation EclipseTarget Masking PointBeamlet Dose at point - d i, Beamlet Beamlet - x j VARIAN ONCOLOGY SYSTEMS
22 IMRT Penalty Calculation A C B 2 1 p1 > 40 w1 = 100 p2 < 20 w2 = 50 A C B 1 2 Next iteration A B C 1 2 d (d violate? penalty Σ A B C 1 2 d (d violate? penalty Σ Gain VARIAN ONCOLOGY SYSTEMS
23 IMRT Objective function ()! F obj = F PTV + F Critical + F NTO + F SmoothXY! allptvs allcs Fields! () + F Min.Dose Fields VARIAN ONCOLOGY SYSTEMS
24 Intermediate Dose Calculation MRDC VARIAN ONCOLOGY SYSTEMS
25 Segment Leaf Motion Calculator optimal fluenceactual fluence LMC Sliding Window Multiple Static Segments None Field size definition VARIAN ONCOLOGY SYSTEMS
26 Segment Large Field IMRT スプリットフィールド ( フィールド分割 ) 幅の広い照射野を左右二つの分割された照射野に変更をする その際 インテンシティーのつながりを考慮して分割する Paul Keall et. al. Virginia Commonwealth University Medical College of Virginia Hospital VARIAN VARIAN ONCOLOGY SYSTEMS
27 Segment Large Field IMRT LargeIMRT VARIAN ONCOLOGY SYSTEMS
28 RapidArc Contouring Auto adjust Field Setup Machine Parameters Optimization Leaf Motions MU per deg MRDC Objectives VARIAN ONCOLOGY SYSTEMS
29 RapidArc Leaf Motions MU per deg AAA or Acuros XB 3D dose calculation Delivery VARIAN ONCOLOGY SYSTEMS
30 VMAT Create Couch Structure Couch structures may be added to account for attenuation of the treatment couch. Numerous Varian couches are available Rails can be positioned in or out, for both left and right Couch position may be adjusted VARIAN ONCOLOGY SYSTEMS
31 VMAT Constraints - Varian Delivery Platforms RT Administration Dose rate / 0.1 MU/deg VARIAN ONCOLOGY SYSTEMS
32 VMAT Geometry 30 Avoidance sectors VMAT Field size Collimator, Couch rotation Isocenter Arc geometry tool VARIAN ONCOLOGY SYSTEMS
33 VMAT Avoidance Sectors VARIAN ONCOLOGY SYSTEMS
34 VMAT RapidArc - Arc Optimization Progressive Resolution Optimization (PRO) Starts with few directions (control points) Optimizes the MLC motion, Dose Rate, Gantry Speed for each control point Respects MLC, Gantry Speed Constraints () VARIAN ONCOLOGY SYSTEMS
35 VMAT Progressive resolution () Control Points resolution multi-resolution Calculation direction for arc segment VARIAN ONCOLOGY SYSTEMS
36 VMAT Steps PRO3 resolution level VARIAN ONCOLOGY SYSTEMS
37 (AcurosXB) Monte Calo f(x,v,t)x,vf(x,v,t)dxdv dxdv!"!" +!!"!" +!!!"!" =!!!!!!!!"Ω!!! v v VARIAN ONCOLOGY SYSTEMS
38 (AcurosXB) Monte Calo VARIAN ONCOLOGY SYSTEMS
39 (AcurosXB) Monte Calo VARIAN ONCOLOGY SYSTEMS
40 (AcurosXB) Monte Calo q γ prim q γ sec q e cont Initial Phase Space q γγ unc q γe unc q ee unc Modified Phase Space VARIAN ONCOLOGY SYSTEMS
41 (AcurosXB) Monte Calo VARIAN ONCOLOGY SYSTEMS
42 より正確な線量計算(AcurosXB) Monte Calo等価な線量計算 不均質層での結果 Monte Carlo計算と同等の結果 5cm 2cm 8cm 6MV, 5x5cm 4cm 11cm 5cm(Water) 8cm(Lung) 2cm(Titanium) VARIAN ONCOLOGY SYSTEMS 11cm(Water) 4cm(Bone)
43 IMRT VARIAN ONCOLOGY SYSTEMS
44 2 VARIAN ONCOLOGY SYSTEMS
45 3 VARIAN ONCOLOGY SYSTEMS
46 4 VARIAN ONCOLOGY SYSTEMS
47 270 3 MLC COPYRIGHT 2010 VARIAN MEDICAL SYSTEMS
48 Clinac Beam Stand Gantry
49 Clinac Beam 4 7 VARIAN ONCOLOGY SYSTEMS Stand Gantry
50 8 VARIAN ONCOLOGY SYSTEMS
51 9 VARIAN ONCOLOGY SYSTEMS
52 10 VARIAN ONCOLOGY SYSTEMS
53
54 3/2
55 0 0 0
56
57 15 VARIAN ONCOLOGY SYSTEMS
58 16 VARIAN ONCOLOGY SYSTEMS
59 x E E x =E 0 sin { (z / c t)} H y =H 0 sin { (z / c t)} z E 0 /H 0 = / (376.7 c=1 / y H EH S = E H z E x =E 0 sin { (z / c + t)} B y = B 0 sin { (z / c + t)} 17 VARIAN ONCOLOGY SYSTEMS
60 TM 010 Ez 2b z 2b E 0 Ez d r Ez z r =b 0 d V=E 0 d/2 d/2 d/2 d/2 cos t dz =E 0 cos z/v+ t 0 dz =V 0 T cos t 0 t=t 0 z=0 V 0 =E 0 d T =sin d d <1 2v 2v T 18 VARIAN ONCOLOGY SYSTEMS
61 1/2 (1) (2) (3) (4) (1) 19 VARIAN ONCOLOGY SYSTEMS
62 2/2 (5) (2) (7) (8) (1) Z 20 VARIAN ONCOLOGY SYSTEMS
63 1 1 1/4/2 /2 1/32/3 2/3 21 VARIAN ONCOLOGY SYSTEMS
64
65 A B C D 2b z d
66 /2 g t=0 t= 1 4 T 0 t= 1 2 T 0 T 0 1/f A B C D
67 /2 0Z
68
69 1 Z=E 02 P 2 ZT 2 (E 0 T) 2 P 2 E 0 T P 27 VARIAN ONCOLOGY SYSTEMS
70 2 (1) (2) T 5060 M/m 100 M/m 28 VARIAN ONCOLOGY SYSTEMS
71 29 VARIAN ONCOLOGY SYSTEMS
72 1 10MeV 6MeV 4MeV 1 V= 1+ 2 LPZT 2 ilzt 2 L m ZT 2 /m P 1 P 2 P 3 A 30 VARIAN ONCOLOGY SYSTEMS
73 2 Buncher Section RF Regular Section ~0.4m ~1m 31 VARIAN ONCOLOGY SYSTEMS
74 2 Buncher Section RF Regular Section ~0.4m ~1m 32 VARIAN ONCOLOGY SYSTEMS
75 33 VARIAN ONCOLOGY SYSTEMS
76 Achromatic Bending Magnet System 34 VARIAN ONCOLOGY SYSTEMS
77 270 C-Series HE Clinac X
78 270 TrueBeam X
79 e 2a B 2b 2d z 37 VARIAN ONCOLOGY SYSTEMS
80 B F = e B B 2 = B(z) d z 1 dz 1 P / e 1 = B(z)dz P / e - 8 = P / e 1 2 abni a 2 +d 2 e z 1 z 2 B z E = 10 (MeV) a = 0.1(m)b = 0.1(m) d = 0.1(m)nI = 100(A) B = P / e = 0.035(Tm) B y (0) = (T) 2.3mrad 38 VARIAN ONCOLOGY SYSTEMS
81
82 40 VARIAN ONCOLOGY SYSTEMS
untitled
SPring-8 RFgun JASRI/SPring-8 6..7 Contents.. 3.. 5. 6. 7. 8. . 3 cavity γ E A = er 3 πε γ vb r B = v E c r c A B A ( ) F = e E + v B A A A A B dp e( v B+ E) = = m d dt dt ( γ v) dv e ( ) dt v B E v E
XFEL/SPring-8
DEVELOPMENT STATUS OF RF SYSTEM OF INJECTOR SECTION FOR XFEL/SPRING-8 Takao Asaka 1,A), Takahiro Inagaki B), Hiroyasu Ego A), Toshiaki Kobayashi A), Kazuaki Togawa B), Shinsuke Suzuki A), Yuji Otake B),
Introduction Purpose This training course demonstrates the use of the High-performance Embedded Workshop (HEW), a key tool for developing software for
Introduction Purpose This training course demonstrates the use of the High-performance Embedded Workshop (HEW), a key tool for developing software for embedded systems that use microcontrollers (MCUs)
q π =0 Ez,t =ε σ {e ikz ωt e ikz ωt } i/ = ε σ sinkz ωt 5.6 x σ σ *105 q π =1 Ez,t = 1 ε σ + ε π {e ikz ωt e ikz ωt } i/ = 1 ε σ + ε π sinkz ωt 5.7 σ
H k r,t= η 5 Stokes X k, k, ε, ε σ π X Stokes 5.1 5.1.1 Maxwell H = A A *10 A = 1 c A t 5.1 A kη r,t=ε η e ik r ωt 5. k ω ε η k η = σ, π ε σ, ε π σ π A k r,t= q η A kη r,t+qηa kηr,t 5.3 η q η E = 1 c A
2
L C -24K 9 L C -22K 9 2 3 4 5 6 7 8 9 10 11 12 11 03 AM 04 05 0 PM 1 06 1 PM 07 00 00 08 2 PM 00 4 PM 011 011 021 041 061 081 051 071 1 2 4 6 8 5 7 00 00 00 00 00 00 00 00 30 00 09 00 15 10 3 PM 45 00
2
WJ-HD150 Digital Disk Recorder WJ-HD150 2 3 q w e r t y u 4 5 6 7 8 9 10 11 12 13 14 15 16 q w SIGNAL GND AC IN 17 SUNDAY MONDAY TUESDAY WEDNESDAY THURSDAY FRIDAY SATURDAY DAILY Program 1 Event No.1 Event
Introduction Purpose This training course describes the configuration and session features of the High-performance Embedded Workshop (HEW), a key tool
Introduction Purpose This training course describes the configuration and session features of the High-performance Embedded Workshop (HEW), a key tool for developing software for embedded systems that
1 2 3
INFORMATION FOR THE USER DRILL SELECTION CHART CARBIDE DRILLS NEXUS DRILLS DIAMOND DRILLS VP-GOLD DRILLS TDXL DRILLS EX-GOLD DRILLS V-GOLD DRILLS STEEL FRAME DRILLS HARD DRILLS V-SELECT DRILLS SPECIAL
ProVisionaire Control V3.0セットアップガイド
ProVisionaire Control V3 1 Manual Development Group 2018 Yamaha Corporation JA 2 3 4 5 NOTE 6 7 8 9 q w e r t r t y u y q w u e 10 3. NOTE 1. 2. 11 4. NOTE 5. Tips 12 2. 1. 13 3. 4. Tips 14 5. 1. 2. 3.
L C -6D Z3 L C -0D Z3 3 4 5 6 7 8 9 10 11 1 13 14 15 16 17 OIL CLINIC BAR 18 19 POWER TIMER SENSOR 0 3 1 3 1 POWER TIMER SENSOR 3 4 1 POWER TIMER SENSOR 5 11 00 6 7 1 3 4 5 8 9 30 1 3 31 1 3 1 011 1
PowerPoint Presentation
2010 KEK (Japan) (Japan) (Japan) Cheoun, Myun -ki Soongsil (Korea) Ryu,, Chung-Yoe Soongsil (Korea) 1. S.Reddy, M.Prakash and J.M. Lattimer, P.R.D58 #013009 (1998) Magnetar : ~ 10 15 G ~ 10 17 19 G (?)
LSM5Pascal Ver 3.2 GFP 4D Image VisArt Carl Zeiss Co.,Ltd.
LSM5Pascal Ver 3.2 GFP 4D Image VisArt 2004.03 LSM5PASCAL V3.2 LSM5PASCAL SW3.2Axiovert200M 1 1 2 3 3 4 4 5 SingleTrack 9 Multi Track 10,18 5 / 21 6 3 27 7 35 8 ( OFF) 40 LSM5PASCAL V3.2 LSM5PASCAL 65
A Feasibility Study of Direct-Mapping-Type Parallel Processing Method to Solve Linear Equations in Load Flow Calculations Hiroaki Inayoshi, Non-member
A Feasibility Study of Direct-Mapping-Type Parallel Processing Method to Solve Linear Equations in Load Flow Calculations Hiroaki Inayoshi, Non-member (University of Tsukuba), Yasuharu Ohsawa, Member (Kobe
Microsoft PowerPoint UM.ppt [互換モード]
CyberKnife の吸収線量測定及び 当院における QA QC への取り組み 横浜サイバーナイフセンター 1) 横浜市立大学大学院 2) 井上光広 1)2) 大川浩平 1) 仙田学 1) 帯刀光史 1) 佐藤健吾 1) 横浜市立大学附属病院小池泉 首都大学東京大学院河内徹 CyberKnife の吸収線量測定 60 Coγ 線以外の線質での水吸収線量 D = M N k w, Q Q D,w,Q
LC304_manual.ai
Stick Type Electronic Calculator English INDEX Stick Type Electronic Calculator Instruction manual INDEX Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union
BRC-X1000
C-455-100-02(1) BRC-X1000/H800 2016 Sony Corporation 2 3 4 5 6 4 D E 5 6 7 A B C F G 7 8 9 0 qa qs SYSTEM SELECT IN IR SELECT VISCA RS - 422 OUT OSD OFF CAMERA SETUP 1 qd 2 3 4 5 6 7 8 LAN 12V qf HDMI
Development of Induction and Exhaust Systems for Third-Era Honda Formula One Engines Induction and exhaust systems determine the amount of air intake
Development of Induction and Exhaust Systems for Third-Era Honda Formula One Engines Induction and exhaust systems determine the amount of air intake supplied to the engine, and as such are critical elements
教育講演 放射線治療における位置的不確かさの影響 ずれるとどうなる 都島放射線科クリニック / 大阪大学塩見浩也 放射線治療を確実に施行するためには, 安全なマージンの設定が不可欠である. ターゲットの設定は,ICRU report 50, 62 3) で規定されている肉眼的腫瘍体積 (GTV; g
教育講演 放射線治療における位置的不確かさの影響 ずれるとどうなる 都島放射線科クリニック / 大阪大学塩見浩也 放射線治療を確実に施行するためには, 安全なマージンの設定が不可欠である. ターゲットの設定は,ICRU report 50, 62 3) で規定されている肉眼的腫瘍体積 (GTV; gross tumor volume), 臨床標的体積 (CTV; clinical target volume),itv(internal
Microsoft PowerPoint - 島田美帆.ppt
コンパクト ERL におけるバンチ圧縮の可能性に関して 分子科学研究所,UVSOR 島田美帆日本原子力研究開発機構,JAEA 羽島良一 Outline Beam dynamics studies for the 5 GeV ERL 規格化エミッタンス 0.1 mm mrad を維持する周回部の設計 Towards user experiment at the compact ERL Short bunch
Accuracy check of grading of XCT Report Accuracy check of grading and calibration of CT value on the micro-focus XCT system Tetsuro Hirono Masahiro Ni
JAMSTEC Rep. Res. Dev., Volume 8, November 2008, 29 36 X CTm/pixel X CT X CT. -. mol/l KI KI CT CT X CT CT ; - - +- -- [email protected] Accuracy check of grading of XCT Report Accuracy check
2
WV-CW180,WV-CW190 WV-CW190 2 q w e r t y 3 4 5 6 7 8 t q y u e!5 w i o!0!4 t q y r!7 @1 @0!9!8!6!1!2!3 e @2 q w e r t y u i o!0!1!2!3!4!5!6!7 @1!7!8!9 @0 @1 @2 9 10 11 12 46 mm 22 mm 83.5 mm 83.5 mm 46
4/15 No.
4/15 No. 1 4/15 No. 4/15 No. 3 Particle of mass m moving in a potential V(r) V(r) m i ψ t = m ψ(r,t)+v(r)ψ(r,t) ψ(r,t) = ϕ(r)e iωt ψ(r,t) Wave function steady state m ϕ(r)+v(r)ϕ(r) = εϕ(r) Eigenvalue problem
Unidirectional Measurement Current-Shunt Monitor with Dual Comparators (Rev. B
www.tij.co.jp INA206 INA207 INA208 INA206-INA208 INA206-INA208 V S 1 14 V IN+ V S 1 10 V IN+ OUT CMP1 IN /0.6V REF 2 3 1.2V REF 13 12 V IN 1.2V REF OUT OUT CMP1 IN+ 2 3 9 8 V IN CMP1 OUT CMP1 IN+ 4 11
LT 低コスト、シャットダウン機能付き デュアルおよびトリプル300MHz 電流帰還アンプ
µ µ LT1398/LT1399 V IN A R G 00Ω CHANNEL A SELECT EN A R F 3Ω B C 97.6Ω CABLE V IN B R G 00Ω EN B R F 3Ω 97.6Ω V OUT OUTPUT (00mV/DIV) EN C V IN C 97.6Ω R G 00Ω R F 3Ω 1399 TA01 R F = R G = 30Ω f = 30MHz
JOURNAL OF THE JAPANESE ASSOCIATION FOR PETROLEUM TECHNOLOGY VOL. 66, NO. 6 (Nov., 2001) (Received August 10, 2001; accepted November 9, 2001) Alterna
JOURNAL OF THE JAPANESE ASSOCIATION FOR PETROLEUM TECHNOLOGY VOL. 66, NO. 6 (Nov., 2001) (Received August 10, 2001; accepted November 9, 2001) Alternative approach using the Monte Carlo simulation to evaluate
cms.pdf
RoHS compliant INTERNAL STRUTURE FEATURES Part name over Slider Housing Slider contact Fixed contact Terminal pin lick spring Ground terminal Material Steel (SP), Tin-plated Polyamide opper alloy, Gold-plated
CMP Technical Report No. 4 Department of Computational Nanomaterials Design ISIR, Osaka University 2 2................................. 2.2......................... 2 3 3 3................................
(e ) (µ ) (τ ) ( (ν e,e ) e- (ν µ,µ ) µ- (ν τ,τ ) τ- ) ( ) ( ) ( ) (SU(2) ) (W +,Z 0,W ) * 1) [ ] [ ] [ ] ν e ν µ ν τ e µ τ, e R,µ R,τ R (2.1a
1 2 2.1 (e ) (µ ) (τ ) ( (ν e,e ) e- (ν µ,µ ) µ- (ν τ,τ ) τ- ) ( ) ( ) ( ) (SU(2) ) (W +,Z 0,W ) * 1) [ ] [ ] [ ] ν e ν µ ν τ e µ τ, e R,µ R,τ R (2.1a) L ( ) ) * 2) W Z 1/2 ( - ) d u + e + ν e 1 1 0 0
main.dvi
A 1/4 1 1/ 1/1 1 9 6 (Vergence) (Convergence) (Divergence) ( ) ( ) 97 1) S. Fukushima, M. Takahashi, and H. Yoshikawa: A STUDY ON VR-BASED MUTUAL ADAPTIVE CAI SYSTEM FOR NUCLEAR POWER PLANT, Proc. of FIFTH
1 Fig. 1 Extraction of motion,.,,, 4,,, 3., 1, 2. 2.,. CHLAC,. 2.1,. (256 ).,., CHLAC. CHLAC, HLAC. 2.3 (HLAC ) r,.,. HLAC. N. 2 HLAC Fig. 2
CHLAC 1 2 3 3,. (CHLAC), 1).,.,, CHLAC,.,. Suspicious Behavior Detection based on CHLAC Method Hideaki Imanishi, 1 Toyohiro Hayashi, 2 Shuichi Enokida 3 and Toshiaki Ejima 3 We have proposed a method for
DICOM Conformance Statement Carino
Carino 2016,03,17 E1J-HC0010-01 Copyright Hitachi, Ltd. 2016. All rights reserved. 1 1 1 2 3 3 AE 3 3 AE 4 4 4 4 4 4 5 5 5 5 SOP SOP 6 6 6 TCP/IP 6 API 6 6 7 SOP 7 7 7 AE 7 7 7 A 8 ( 2 ) E1J-HC0010 DICOM
2
L C -60W 7 2 3 4 5 6 7 8 9 0 2 3 OIL CLINIC BAR 4 5 6 7 8 9 2 3 20 2 2 XXXX 2 2 22 23 2 3 4 5 2 2 24 2 2 25 2 3 26 2 3 6 0 2 3 4 5 6 7 8 9 2 3 0 2 02 4 04 6 06 8 08 5 05 2 3 4 27 2 3 4 28 2 3 4 5 2 2
Study on Application of the cos a Method to Neutron Stress Measurement Toshihiko SASAKI*3 and Yukio HIROSE Department of Materials Science and Enginee
Study on Application of the cos a Method to Neutron Stress Measurement Toshihiko SASAKI*3 and Yukio HIROSE Department of Materials Science and Engineering, Kanazawa University, Kakuma-machi, Kanazawa-shi,
PRODUCT INFORMATION Highly Efficient FXS Carbide Ball Nose End Mills Vol. 3 PAT.P. FXS-EBT FXS-LS-EBT FXS-PC-EBT FXS-EBM
PRODUCT INFORMATION Highly Efficient FXS Carbide Ball Nose End Mills Vol. 3 PAT.P. FXS-EBT FXS-LS-EBT FXS-PC-EBT FXS-EBM 3 Flutes Series Features Thanks to 3 flutes ball nose geometry, all of that reach
06_学術_関節単純X線画像における_1c_梅木様.indd
Arts and Sciences X The formulation of femoral heard measurement corrected enlargement ratio using hip joints X-ray Imaging 1 2 1 1 1 2 Key words: Bipolar Hip Arthroplasty (BHA) Preoperative planning Enlargement
EGunGPU
Super Computing in Accelerator simulations - Electron Gun simulation using GPGPU - K. Ohmi, KEK-Accel Accelerator Physics seminar 2009.11.19 Super computers in KEK HITACHI SR11000 POWER5 16 24GB 16 134GFlops,
AX-V8000[DOM]
5 3 7 8 SPEAKERS SPEAKERS 2 INPUT MODE INPUT ATT SURROUND ON/OFF HTX ON / OFF DSP MODE LINE DIRECT SOUND SELECTOR REC SELECTOR ADJUST MENU SETUP MENU AUDIO/VIDEO CONTROL AMPLIFIER DOWN LEFT UP RIGHT SET
DOI: /jjmu. JJMU.R.130 REVIEW ARTICLE( 第 16 回教育セッション )( 乳腺 ) 抄録,, 1 surrounding stroma 2 edematous stroma 2., TDLU,,.,.,,.,.,,.,.,,,,,. Ultraso
DOI: 10.3179/jjmu. JJMU.R.130 REVIEW ARTICLE( 第 16 回教育セッション )( 乳腺 ) 抄録, 1 surrounding stroma 2 edematous stroma 2, TDLU,,,.,,,,,, Ultrasound understanding and observation method of normal mammary gland
K02LE indd
I Linear Stage Table of Contents 1. Features... 1 2. Description of part Number... 1 3. Maximum Speed Limit... 2 4. Specifications... 3 5. Accuracy Grade... 4 6. Motor and Motor Adaptor Flange... 4 6-1
スペースプラズマ研究会-赤星.ppt
14 1 1 1 1 Pauline Faure 1 1 2 3 (1: 2: JAXA 3: IHI) IHI (C)(No.21560819) ISAS(JAXA) ISO TC20/SC14 / (Spall) 60~90% 2 (Cone) 1% (Jetting) CDV11227 Committee Draft for Comments CDV11227 Witness plate Sabot
GP05取説.indb
E -G V P 05D L V E -G P 05D W Ni-MH + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + 1 + 2 + 3 + 4 + 5 + 6 1 2 3 4 5 6 + + + 1 + + + + + + + + + + + + + + + + + + 1 A B C + D + E
T rank A max{rank Q[R Q, J] t-rank T [R T, C \ J] J C} 2 ([1, p.138, Theorem 4.2.5]) A = ( ) Q rank A = min{ρ(j) γ(j) J J C} C, (5) ρ(j) = rank Q[R Q,
(ver. 4:. 2005-07-27) 1 1.1 (mixed matrix) (layered mixed matrix, LM-matrix) m n A = Q T (2m) (m n) ( ) ( ) Q I m Q à = = (1) T diag [t 1,, t m ] T rank à = m rank A (2) 1.2 [ ] B rank [B C] rank B rank
DL1620/DL1640/DL1640L ディジタルオシロスコープ ユーザーズマニュアル
DL1620/DL1640/DL1640L ディジタルオシロスコープ ユーザーズマニュアル IM 701610-01 8 版 ユーザー登録のお願い 今後の新製品情報を確実にお届けするために お客様にユーザー登録をお願いしております 下記 URL の ユーザー登録 のページで ご登録いただけます http://www.yokogawa.co.jp/tm/ 計測相談のご案内 当社では お客様に正しい計測をしていただけるよう
untitled
January 2009 Rotor-Gene Q Sample & Assay Technologies 1 1-1 1.1 1-1 1.2 Rotor-Gene Q 1-1 1.3 Rotor-Gene Q 1-1 1.4 1-1 2 2-1 2.1 Run File 2-1 2.2 2-2 2.3 2-2 2.4 PCR 2-3 2.5 2-3 2.6 2-4 2.7 2-4 2.8 2-5
DL1700Eシリーズ ディジタルオシロスコープ ユーザーズマニュアル
DL1700E シリーズ ディジタルオシロスコープ ユーザーズマニュアル IM 701730-01 3 版 ユーザー登録のお願い 今後の新製品情報を確実にお届けするために お客様にユーザー登録をお願いしております 下記 URL の ユーザー登録 のページで ご登録いただけます http://www.yokogawa.co.jp/tm/ 計測相談のご案内 当社では お客様に正しい計測をしていただけるよう
1 1 2 2 3 3 RBS 3 K-factor 3 5 8 Bragg 15 ERDA 21 ERDA 21 ERDA 21 31 31 33 RUMP 38 42 42 42 42 42 42 4 45 45 Ti 45 Ti 61 Ti 63 Ti 67 Ti 84 i Ti 86 V 90 V 99 V 101 V 105 V 114 V 116 121 Ti 121 Ti 123 V
EQUIUM EQUIUM S5010 1
EQUIUM EQUIUM S5010 1 1 1 2 3 4 2 1 2 3 2 3 1 2 3 4 5 6 7 4 1 2 5 1 2 6 1 7 1 2 3 4 5 6 7 1 2 3 4 5 6 3 7 8 9 1 2 3 4 4 5 6 7 8 1 1 2 3 4 10 1 11 12 1 13 14 1 15 16 1 1 2 3 17 1 2 3 18 4 5 6 1 19 1 2 3
Triple 2:1 High-Speed Video Multiplexer (Rev. C
www.tij.co.jp OPA3875 µ ± +5V µ RGB Channel OPA3875 OPA3875 (Patented) RGB Out SELECT ENABLE RED OUT GREEN OUT BLUE OUT 1 R G B RGB Channel 1 R1 G1 B1 X 1 Off Off Off 5V Channel Select EN OPA875 OPA4872
/ Motor Specifications Direct Motor Drive Ball Screws / Precision Ball Screw type MB / MB MB Precision Ball Screw type MB / MoBo C3 5 5 Features A 5-p
/ Motor Specifications MB Precision Ball Screw type MB / MoBo C3 5 5 Features A 5-pahse Stepping Motor is mounted directly onto the shaft end of a C3 grade precision Ball Screw, which is suitable for high
アクアEXオイルホールロング(CC2015).indd
High Efficiency Deep Hole Drilling D 1D 20D 2D 0D UA Drills Oil-Hole Long Non-Step drilling is possible at the hole depth of 0D High efficient drilling with long tool life with Wet and MQL Cutting edge
Fig. 3 Flow diagram of image processing. Black rectangle in the photo indicates the processing area (128 x 32 pixels).
Fig. 1 The scheme of glottal area as a function of time Fig. 3 Flow diagram of image processing. Black rectangle in the photo indicates the processing area (128 x 32 pixels). Fig, 4 Parametric representation
Microsoft PowerPoint - 山形大高野send ppt [互換モード]
, 2012 10 SCOPE, 2012 10 2 CDMA OFDMA OFDM SCOPE, 2012 10 OFDM 0-20 Relative Optical Power [db] -40-60 10 Gbps NRZ BPSK-SSB 36dB -80-20 -10 0 10 20 Relative Frequency [GHz] SSB SSB OFDM SSB SSB OFDM OFDM
HT-K215R
4-237-544-03(3) HT-K215R HT-K215M 2001 Sony Corporation z * DTS** * DolbyPro LogicD **Digital Theater Systems,Inc DTSDTS Digital Surround Digital Theater Systems,Inc 2 ...4... 4... 5... 6... 7...9... 9...
報告書
2018 年 1 月 5 日 郵送 IMRT 第三者評価プログラム ~ JCOG 郵送 IMRT 測定 ~ JCOG 放射線治療医学物理ワーキンググループ郵送 IMRT 測定担当者岡本裕之 峯村俊行 ( 連絡先 ) 104-0045 東京都中央区築地 5-1-1 国立がん研究センターがん対策情報センターがん医療支援部放射線治療品質管理推進室峯村俊行 TEL:03-3542-2511 ( 内線 1700)
AX-D701[DOM]
AX-D701 4 7 LVT1438-001A 1 2 47 8 8 9 12 12 14 14 16 18 20 22 22 23 23 24 26 2728 2930 29 29 29 30 30 3133 31 31 344 343 36 37 37 37 38 38 40 40 41 41 42 42 43 44 44 4 4 462 4647 48 49 49 0 0 0 0 1 1 1
The Evaluation on Impact Strength of Structural Elements by Means of Drop Weight Test Elastic Response and Elastic Limit by Hiroshi Maenaka, Member Sh
The Evaluation on Impact Strength of Structural Elements by Means of Drop Weight Test Elastic Response and Elastic Limit by Hiroshi Maenaka, Member Shigeru Kitamura, Member Masaaki Sakuma Genya Aoki, Member
64 3 g=9.85 m/s 2 g=9.791 m/s 2 36, km ( ) 1 () 2 () m/s : : a) b) kg/m kg/m k
63 3 Section 3.1 g 3.1 3.1: : 64 3 g=9.85 m/s 2 g=9.791 m/s 2 36, km ( ) 1 () 2 () 3 9.8 m/s 2 3.2 3.2: : a) b) 5 15 4 1 1. 1 3 14. 1 3 kg/m 3 2 3.3 1 3 5.8 1 3 kg/m 3 3 2.65 1 3 kg/m 3 4 6 m 3.1. 65 5
Postoperative Assessment by Using a Torque Machine (BIODEX) and MRIs in Patients treated for a Recurrent Anterior Dislocation or Subluxation of the Sh
Postoperative Assessment by Using a Torque Machine (BIODEX) and MRIs in Patients treated for a Recurrent Anterior Dislocation or Subluxation of the Shoulder by HATA Yukihiko MURAKAMI Narumichi Department
01.eps
444 SumiTurn T-REX Tool Holders Cost reduction for copying with unique multiple cornered insert Carbide inserts, cermet inserts and 4 types of breakers are available. Expansion of our popular steel turning
GM-F520S/GM-F470S/GM-F420S
GM-F520S GM-F470S GM-F420S LCT2504-002A-H 2 3 4 200 150 150 50 1 3 4 1 2 3 1 2 3 4 5 e 6 7 8 9 p q w r t 5 6 5 23 7 8 9 p q 4 5 6 7 8 9 2 3 4 5 5 23 6 7 8 9 w 1 2 e r t p p 5 6 9( 3 DVI-D (HDCP) RGB IN
25 3 4
25 3 4 1 µ e + ν e +ν µ µ + e + +ν e + ν µ e e + TAC START STOP START veto START (2.04 ± 0.18)µs 1/2 STOP (2.09 ± 0.11)µs 1/8 G F /( c) 3 (1.21±0.09) 5 /GeV 2 (1.19±0.05) 5 /GeV 2 Weinberg θ W sin θ W
Monaco コンベンショナルトレーニング へようこそ。
Monaco Training Conventional Course Monaco5.11 エレクタ株式会社 目次 計画ツール 4 新規プランの作成処方線量の入力ビームの操作ビームスプレッドシート Beam Visibility ポートの描画ポートの自動適合ポートまたは MLC の編集 MLC が編集できない 計算プロパティフルエンスマップ 計画レビュー 28 正規化パラメータ等線量表示の変更 Template
http://www.ns.kogakuin.ac.jp/~ft13389/lecture/physics1a2b/ pdf I 1 1 1.1 ( ) 1. 30 m µm 2. 20 cm km 3. 10 m 2 cm 2 4. 5 cm 3 km 3 5. 1 6. 1 7. 1 1.2 ( ) 1. 1 m + 10 cm 2. 1 hr + 6400 sec 3. 3.0 10 5 kg
5 1 2 3 4 5 6 7 8 9 10 1 Quick Boot 1st Boot Device 2nd Boot Device 3rd Boot Device Network Boot Initial Display Mode S.M.A.R.T. for Hard Disks BootUp Num-Lock Password Check CPU Serial Number System BIOS
1..FEM FEM 3. 4.
008 stress behavior at the joint of stringer to cross beam of the steel railway bridge 1115117 1..FEM FEM 3. 4. ABSTRACT 1. BackgroundPurpose The occurrence of fatigue crack is reported in the joint of
a a b a b c d e R c d e A a b e a b a b c d a b c d e f a M a b f d a M b a b a M b a M b M M M R M a M b M c a M a R b A a b b a CF a b c a b a M b a b M a M b c a A b a b M b a A b a M b C a M C a M
LC-24_22_19K30.indb
L C -24K 30 L C -22K 30 L C -19K 30 http://www.sharp.co.jp/support/aquos/ 2 3 4 5 6 7 8 LC-24K30 9 10 11 12 LC-24K30 8 10 PM 11 12 9 PM 13 10 PM 14 11 15 PM 16 0 17 AM 1 3 101 103 00 00 30 50 00 00 00
Microsoft Word - 01_巻頭言.docx
テーマ 肺がん治療の臨床現場での取り組み 肺がん放射線治療への取り組み - 呼吸性移動対策と FFF の臨床利用 - 神戸低侵襲がん医療センター上原和之 はじめに当院では肺がんの治療に対して CyberKnife(Accuray 社製 ) と TrueBeam(Varian Medical Systems 社製 ) の 2 つの放射線治療装置を運用し, 症例によって選択している. 今回は, 放射線治療装置の使い分けと実施している呼吸性移動対策,Flattening
Table 1: Basic parameter set. Aperture values indicate the radius. δ is relative momentum deviation. Parameter Value Unit Initial emittance 10 mm.mrad
SuperKEKB EMITTANCE GROWTH BY MISALIGNMENTS AND JITTERS IN SUPERKEKB INJECTOR LINAC Y. Seimiya, M. Satoh, T. Suwada, T. Higo, Y. Enomoto, F. Miyahara, K. Furukawa High Energy Accelerator Research Organization
